Comparisons of Two Bioelectrical Impedance Devices and Manual versus Sensor-Based Short Physical Performance Batteries for Assessment of Muscle Mass and Physical Performance

Jin-Young Min iD, Kyoung-Bok Min

Sensors · 2023

DOI: 10.3390/s23136026 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2023. Verfasst von Jin-Young Min und 1 weiteren Autor:innen. Beteiligte Einrichtung: Veterans Medical Research Institute, Veterans Health Service Medical Center, Seoul 05368, Republic of Korea. Hauptthemen laut MEDLINE: Muscle, Skeletal, Sarcopenia. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: VHS Medical Center Research Grant, Republic of Korea.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The assessment of muscle mass and physical performance is essential for the diagnosis of sarcopenia. This study examined the validity of bioimpedance analysis (BIA) and a sensor-based short physical performance battery (SPPB) device for analyzing appendicular skeletal muscle mass (ASM) and physical performance. Forty older adults were measured for ASM and physical performance with two BIA devices (InBody770 vs. T-SCAN PLUS III) and two SPPB devices (manual mSPPB vs. sensor-based sSPPB). Validity statistics included the intraclass correlation coefficient (ICC) and Bland–Altman plots to examine the agreement of data from the BIA (InBody770 vs. T-SCAN PLUS III) and the SPPBs (mSPPB vs. sSPPB). There was a significant ICC for skeletal muscle mass between the T-SCAN PLUS III and InBody770 devices (ICC = 0.8822; p < 0.0001). The mSPPB and sSPPB values showed agreement across all components: 0.8654 for the total scores, 0.8879 for the walking speed, 0.8889 for the chair stand, and 0.6863 for the standing balance. No systemic bias was observed between the two methods for the BIA and SPPB devices. Measurements using the T-SCAN PLUS III and sSPPB seem to be highly correlated with the InBody770 and mSPPB devices in older adults and may be valid for assessing muscle mass and physical performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Muscle, SkeletalSarcopeniaHumansAgedBody Mass IndexElectric ImpedancePhysical Functional PerformanceBody CompositionAbsorptiometry, Photon

Stichwörter der Autor:innen

bioimpedance analysismuscle massphysical functionsarcopeniavalidation

Autor:innen und Institutionen

  • Veterans Medical Research Institute, Veterans Health Service Medical Center, Seoul 05368, Republic of Korea.
  • Department of Preventive Medicine, College of Medicine, Seoul National University, Seoul 08826, Republic of Korea.
  • Institute of Health Policy and Management, Medical Research Center, Seoul National University, Seoul 03080, Republic of Korea.

ORCID-Profile: Jin-Young Min

Förderung

  • VHS Medical Center Research Grant, Republic of Korea

Zitieren

APA

Min, J. Y., & Min, K. B. (2023). Comparisons of Two Bioelectrical Impedance Devices and Manual versus Sensor-Based Short Physical Performance Batteries for Assessment of Muscle Mass and Physical Performance. Sensors. https://doi.org/10.3390/s23136026

Vancouver

Min JY, Min KB. Comparisons of Two Bioelectrical Impedance Devices and Manual versus Sensor-Based Short Physical Performance Batteries for Assessment of Muscle Mass and Physical Performance. Sensors. 2023. doi:10.3390/s23136026

BibTeX
@article{Min2023,
  author = {Min, Jin-Young and Min, Kyoung-Bok},
  title = {Comparisons of Two Bioelectrical Impedance Devices and Manual versus Sensor-Based Short Physical Performance Batteries for Assessment of Muscle Mass and Physical Performance},
  journal = {Sensors},
  year = {2023},
  doi = {10.3390/s23136026}
}
Kostenlos starten →