The Validity of Hawkin Dynamics Wireless Dual Force Plates for Measuring Countermovement Jump and Drop Jump Variables

Andrew J Badby, Peter D Mundy, Paul Comfort iD, Jason P Lake iD, John J McMahon iD

Sensors · 17.05.2023 · 23 (10) · 4820

DOI: 10.3390/s23104820 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, online erschienen am 17.05.2023. Verfasst von Andrew J Badby und 4 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Human Movement and Rehabilitation, University of Salford, Salford M6 6PU, UK. Die Arbeit wurde bisher 109-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Hawkin Dynamics Inc. and University of Salford (funded PhD), PhD studentship via Hawkin Dynamics Inc., University of Salford.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Force plate testing is becoming more commonplace in sport due to the advent of commercially available, portable, and affordable force plate systems (i.e., hardware and software). Following the validation of the Hawkin Dynamics Inc. (HD) proprietary software in recent literature, the aim of this study was to determine the concurrent validity of the HD wireless dual force plate hardware for assessing vertical jumps. During a single testing session, the HD force plates were placed directly atop two adjacent Advanced Mechanical Technology Inc. in-ground force plates (the “gold standard”) to simultaneously collect vertical ground reaction forces produced by 20 participants (27 ± 6 years, 85 ± 14 kg, 176.5 ± 9.23 cm) during the countermovement jump (CMJ) and drop jump (DJ) tests (1000 Hz). Agreement between force plate systems was determined via ordinary least products regression using bootstrapped 95% confidence intervals. No bias was present between the two force plate systems for any of the CMJ and DJ variables, except DJ peak braking force (proportional bias) and DJ peak braking power (fixed and proportional bias). The HD system may be considered a valid alternative to the industry gold standard for assessing vertical jumps because fixed or proportional bias was identified for none of the CMJ variables (n = 17) and only 2 out of 18 DJ variables.

Schlagwörter

Stichwörter der Autor:innen

accuracyagreementconcurrent validityforce–timehardwarevertical jumpwireless dual force plate system

Autor:innen und Institutionen

  • Centre for Human Movement and Rehabilitation, University of Salford, Salford M6 6PU, UK.
  • Hawkin Dynamics, Inc., Westbrook, ME 04092, USA.
  • School of Medical and Health Sciences, Edith Cowan University, Joondalup, WA 6027, Australia.
  • Chichester Institute of Sport, University of Chichester, Chichester PO19 6PE, UK.

ORCID-Profile: Paul Comfort, Jason P Lake, John J McMahon

Förderung

  • Hawkin Dynamics Inc. and University of Salford (funded PhD)
  • PhD studentship via Hawkin Dynamics Inc.
  • University of Salford

Zitieren

APA

Badby, A. J., Mundy, P. D., Comfort, P., Lake, J. P., & McMahon, J. J. (2023). The Validity of Hawkin Dynamics Wireless Dual Force Plates for Measuring Countermovement Jump and Drop Jump Variables. Sensors, 23(10), 4820. https://doi.org/10.3390/s23104820

Vancouver

Badby AJ, Mundy PD, Comfort P, Lake JP, McMahon JJ. The Validity of Hawkin Dynamics Wireless Dual Force Plates for Measuring Countermovement Jump and Drop Jump Variables. Sensors. 2023;23(10):4820. doi:10.3390/s23104820

BibTeX
@article{Badby2023,
  author = {Badby, Andrew J and Mundy, Peter D and Comfort, Paul and Lake, Jason P and McMahon, John J},
  title = {The Validity of Hawkin Dynamics Wireless Dual Force Plates for Measuring Countermovement Jump and Drop Jump Variables},
  journal = {Sensors},
  year = {2023},
  volume = {23},
  number = {10},
  pages = {4820},
  doi = {10.3390/s23104820}
}
Kostenlos starten →