In Situ Calibration of Wetlabs Chlorophyll Sensors: A Methodology Adapted to Profile Measurements

Joëlle Salaün, Marc Le Menn iD

Sensors · 2023

DOI: 10.3390/s23052825 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2023. Verfasst von Joëlle Salaün und 1 weiteren Autor:innen. Beteiligte Einrichtung: Shom, French Navy Hydrographic and Oceanographic Service, 29228 Brest, France. Hauptthemen laut MEDLINE: Chlorophyll, Environmental Monitoring. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Measurement of chlorophyll a content in the ocean is essential for biomass assessment, finding the optical properties of seawater, and calibration of satellite remote sensing. The instruments used for this purpose are mostly fluorescence sensors. The calibration of these sensors becomes a crucial point to ensure the reliability and quality of the data produced. The technology of these sensors is based on the principle that a concentration of chlorophyll a in µg per liter can be calculated from an in situ fluorescence measurement. However, the study of the phenomenon of photosynthesis and cell physiology teaches us that the yield of fluorescence depends on many factors that are difficult or impossible to reconstitute in a metrology laboratory. This is the case, for example, of the algal species, its physiological state, the greater or lesser presence of dissolved organic matter in the water, the turbidity of the environment, or the surface illumination. What approach should be adopted in this context to achieve a better quality of the measurements? This is the objective of the work we present here, which is the result of nearly ten years of experimentation and testing to optimize the metrological quality of chlorophyll a profile measurement. The results we obtained allowed us to calibrate these instruments with an uncertainty of 0.2–0.3 on the correction factor, with correlation coefficients higher than 0.95 between the sensor values and the reference value.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ChlorophyllEnvironmental MonitoringChlorophyll ACalibrationReproducibility of Results

Stichwörter der Autor:innen

HPLCWetlabscalibrationchlorophyll sensorin situ measurements

Substanzen

ChlorophyllChlorophyll A

Autor:innen und Institutionen

  • Shom, French Navy Hydrographic and Oceanographic Service, 29228 Brest, France.

ORCID-Profile: Marc Le Menn

Zitieren

APA

Salaün, J., & Menn, M. L. (2023). In Situ Calibration of Wetlabs Chlorophyll Sensors: A Methodology Adapted to Profile Measurements. Sensors. https://doi.org/10.3390/s23052825

Vancouver

Salaün J, Menn ML. In Situ Calibration of Wetlabs Chlorophyll Sensors: A Methodology Adapted to Profile Measurements. Sensors. 2023. doi:10.3390/s23052825

BibTeX
@article{Salan2023,
  author = {Salaün, Joëlle and Menn, Marc Le},
  title = {In Situ Calibration of Wetlabs Chlorophyll Sensors: A Methodology Adapted to Profile Measurements},
  journal = {Sensors},
  year = {2023},
  doi = {10.3390/s23052825}
}
Kostenlos starten →