High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical Devices
Sensors · 2023
DOI: 10.3390/s23031209 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2023. Verfasst von Nandita Saha und 7 weiteren Autor:innen. Beteiligte Einrichtung: Max-Delbrück-Center for Molecular Medicine in the Helmholtz Association (MDC), Berlin Ultrahigh Field Facility (B.U.F.F.), 13125 Berlin, Germany. Hauptthemen laut MEDLINE: Electromagnetic Fields, Defibrillators, Implantable. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Research Council (ERC) under the European Union's Horizon 2020 Research and In-novation Program, European Horizon 2020 MSCA-Innovative Training Network, European Research Council (ERC).
Abstract (Originalsprache)
Case reports indicate that magnets in smartphones could be a source of electromagnetic interference (EMI) for active implantable medical devices (AIMD), which could lead to device malfunction, compromising patient safety. Recognizing this challenge, we implemented a high-fidelity 3D magnetic field mapping (spatial resolution 1 mm) setup using a three-axis Hall probe and teslameter, controlled by a robot (COSI Measure). With this setup, we examined the stray magnetic field of an iPhone 13 Pro, iPhone 12, and MagSafe charger to identify sources of magnetic fields for the accurate risk assessment of potential interferences with AIMDs. Our measurements revealed that the stray fields of the annular array of magnets, the wide-angle camera, and the speaker of the smartphones exceeded the 1 mT limit defined by ISO 14117:2019. Our data-driven safety recommendation is that an iPhone 13 Pro should be kept at least 25 mm away from an AIMD to protect it from unwanted EMI interactions. Our study addresses safety concerns due to potential device–device interactions between smartphones and AIMDs and will help to define data-driven safety guidelines. We encourage vendors of electronic consumer products (ECP) to provide information on the magnetic fields of their products and advocate for the inclusion of smartphones in the risk assessment of EMI with AIMDs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Max-Delbrück-Center for Molecular Medicine in the Helmholtz Association (MDC), Berlin Ultrahigh Field Facility (B.U.F.F.), 13125 Berlin, Germany.
- Experimental and Clinical Research Center (ECRC), A Joint Cooperation between the Charité Medical Faculty and the Max-Delbrück Center for Molecular Medicine in the Helmholtz Association, 13125 Berlin, Germany.
- Department of Physics, Humboldt University of Berlin, 10117 Berlin, Germany.
- Chair of Medical Engineering, Technische Universität Berlin, 10623 Berlin, Germany.
- Department of Cardiology, Charité-Universitätsmedizin Berlin, Campus Virchow-Klinikum, 13353 Berlin, Germany.
ORCID-Profile: Nandita Saha, Jason M Millward, Carl J J Herrmann, Faezeh Rahimi, Thoralf Niendorf
Förderung
- European Research Council (ERC) under the European Union's Horizon 2020 Research and In-novation Program
- European Horizon 2020 MSCA-Innovative Training Network
- European Research Council (ERC)
Zitieren
APA
Saha, N., Millward, J. M., Herrmann, C. J. J., Rahimi, F., Han, H., Lacour, P., Blaschke, F., & Niendorf, T. (2023). High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical Devices. Sensors. https://doi.org/10.3390/s23031209
Vancouver
Saha N, Millward JM, Herrmann CJJ, Rahimi F, Han H, Lacour P, et al. High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical Devices. Sensors. 2023. doi:10.3390/s23031209
BibTeX
@article{Saha2023,
author = {Saha, Nandita and Millward, Jason M and Herrmann, Carl J J and Rahimi, Faezeh and Han, Haopeng and Lacour, Philipp and Blaschke, Florian and Niendorf, Thoralf},
title = {High-Fidelity 3D Stray Magnetic Field Mapping of Smartphones to Address Safety Considerations with Active Implantable Electronic Medical Devices},
journal = {Sensors},
year = {2023},
doi = {10.3390/s23031209}
}
EUVIMED