Patient Radiation Dose During Fluoroscopy-Guided Peripherally Inserted Central Catheter (PICC) Placement

Masakatsu Tano, Kodai Sagehashi, Koichi Chida

Radiation · 10.03.2026 · 6 (1) · 9

DOI: 10.3390/radiation6010009 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation, online erschienen am 10.03.2026. Verfasst von Masakatsu Tano und 2 weiteren Autor:innen. Die Arbeit nennt 68 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Industrial Disease Clinical Research.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This retrospective study evaluated patient radiation dose during fluoroscopy-guided peripherally inserted central catheter (PICC) placement. A total of 1240 consecutive adult patients who underwent PICC placement between January 2023 and December 2024 were analyzed. Patient radiation dose indices, including air kerma (AK) and dose–area product (DAP), as well as fluoroscopy time and number of radiographic acquisitions, were obtained from the radiology information system. The mean and median AK were 2.47 mGy and 1.54 mGy, respectively, and the median DAP was 901.9 mGy·cm2. The median fluoroscopy time was 1.9 min, and the median number of radiographic acquisitions was 1. Patient radiation dose during PICC placement was lower than the Japanese Diagnostic Reference Levels (Japan DRLs 2025). AK showed a strong positive correlation with fluoroscopy time (Spearman’s rank correlation, ρ = 0.77), whereas correlations between AK and BMI or the number of radiographic acquisitions were weak. In some patients with high BMI, AK values exceeding 40 mGy were observed. These findings indicate that patient radiation dose during PICC placement is generally low but remains closely associated with fluoroscopy time. Optimization of the patient radiation dose should be considered, particularly for patients with high BMIs or those undergoing repeated PICC placements.

Förderung

  • Industrial Disease Clinical Research

Zitieren

APA

Tano, M., Sagehashi, K., & Chida, K. (2026). Patient Radiation Dose During Fluoroscopy-Guided Peripherally Inserted Central Catheter (PICC) Placement. Radiation, 6(1), 9. https://doi.org/10.3390/radiation6010009

Vancouver

Tano M, Sagehashi K, Chida K. Patient Radiation Dose During Fluoroscopy-Guided Peripherally Inserted Central Catheter (PICC) Placement. Radiation. 2026;6(1):9. doi:10.3390/radiation6010009

BibTeX
@article{Tano2026,
  author = {Tano, Masakatsu and Sagehashi, Kodai and Chida, Koichi},
  title = {Patient Radiation Dose During Fluoroscopy-Guided Peripherally Inserted Central Catheter (PICC) Placement},
  journal = {Radiation},
  year = {2026},
  volume = {6},
  number = {1},
  pages = {9},
  doi = {10.3390/radiation6010009}
}
Kostenlos starten →