EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Finite Element Analysis of Prosthetic Finger Connectors: Comparison of Ti-6Al-4V and PEEK-Based Polymers Using COMSOL

Nazgul Assan, Kanat Suleimenov, Aiman Ozhikenova, Zhadyra Alimbayeva, Assylbek Ozhiken, Akim Kapsalyamov

Prosthesis · 01.08.2026 · 8 (8) · 81

DOI: 10.3390/prosthesis8080081 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Prosthesis, online erschienen am 01.08.2026. Verfasst von Nazgul Assan und 5 weiteren Autor:innen. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Committee of Science of the Ministry of Science and Higher Education of the Republic of Kazakhstan.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction/Objectives: The mechanical reliability of prosthetic finger connectors is critical for daily prosthesis use, as these small load-bearing elements transmit forces between finger segments and are prone to localized stress concentration at hinge transitions. The objective of this study was to quantitatively evaluate the mechanical performance of prosthetic finger connectors and determine how material selection influences stress distribution and deformation behavior when connector geometry is maintained constant.

Methods

A controlled comparative finite element analysis was performed using COMSOL Multiphysics 6.2 on a single prosthetic finger connector geometry manufactured from four candidate materials: titanium alloy Ti-6Al-4V (E = 110 GPa), polyetheretherketone (PEEK, E = 4.0 GPa), AvaSpire AV-621 NT (E = 3.0 GPa), and KetaSpire KT-820 FP (E = 4.1 GPa). Three representative loading scenarios were simulated using an axial grip force of 50 N and a bending moment of 0.5 N·m: (1) axial loading, (2) combined axial and forward bending, and (3) combined axial and reverse bending.

Results

Across all loading conditions, peak von Mises stresses consistently localized at the hinge-transition region of the connector. Ti-6Al-4V demonstrated the lowest peak stress (approximately 2.2 MPa under combined forward bending) and the smallest maximum displacement (approximately 0.0005 mm). In contrast, the PEEK-family polymers showed greater deformation (0.0125–0.0167 mm) due to their lower elastic modulus, while maintaining comparable peak stresses of approximately 2.5–2.6 MPa. The stress profiles along the connector arc length were nearly identical among all materials, indicating that connector geometry primarily governs stress localization, whereas material stiffness mainly determines deformation magnitude.

Conclusions

This study demonstrates that material selection strongly affects prosthetic finger connector compliance, while stress concentration behavior is mainly controlled by geometry. The obtained quantitative comparison provides design-oriented guidance for selecting between Ti-6Al-4V and PEEK-based polymers for the development of lightweight and mechanically reliable prosthetic finger connectors.

Förderung

  • Committee of Science of the Ministry of Science and Higher Education of the Republic of Kazakhstan

Zitieren

APA

Assan, N., Suleimenov, K., Ozhikenova, A., Alimbayeva, Z., Ozhiken, A., & Kapsalyamov, A. (2026). Finite Element Analysis of Prosthetic Finger Connectors: Comparison of Ti-6Al-4V and PEEK-Based Polymers Using COMSOL. Prosthesis, 8(8), 81. https://doi.org/10.3390/prosthesis8080081

Vancouver

Assan N, Suleimenov K, Ozhikenova A, Alimbayeva Z, Ozhiken A, Kapsalyamov A. Finite Element Analysis of Prosthetic Finger Connectors: Comparison of Ti-6Al-4V and PEEK-Based Polymers Using COMSOL. Prosthesis. 2026;8(8):81. doi:10.3390/prosthesis8080081

BibTeX
@article{Assan2026,
  author = {Assan, Nazgul and Suleimenov, Kanat and Ozhikenova, Aiman and Alimbayeva, Zhadyra and Ozhiken, Assylbek and Kapsalyamov, Akim},
  title = {Finite Element Analysis of Prosthetic Finger Connectors: Comparison of Ti-6Al-4V and PEEK-Based Polymers Using COMSOL},
  journal = {Prosthesis},
  year = {2026},
  volume = {8},
  number = {8},
  pages = {81},
  doi = {10.3390/prosthesis8080081}
}
Kostenlos starten →