Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study
Polymers · 11.06.2026 · 18 (12) · 1457
DOI: 10.3390/polym18121457 ↗Kurz gefasst
Fachartikel in der Zeitschrift Polymers, online erschienen am 11.06.2026. Verfasst von Maria Luís Basto und 5 weiteren Autor:innen. Beteiligte Einrichtung: University of Coimbra, Centre for Mechanical Engineering, Materials and Processes (CEMMPRE) & Department of Dentistry, Faculty of Medicine, 3000-075 …. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Implant-Supported Fixed Prostheses (ISFPs) have become a common option for the rehabilitation of fully edentulous arches and have traditionally incorporated metallic substructures with ceramic or acrylic veneering. The rapid expansion of CAD/CAM technologies has introduced not only a range of polymer-based materials as alternatives to conventional metallic frameworks but also the possibility of the fabrication of monolithic rehabilitations. However, the evidence regarding the mechanical behavior of monolithic polymer-based full-arch rehabilitations remains limited. This study aimed to evaluate and compare the mechanical performance of monolithic polymer-based complete prostheses under static loading using Digital Image Correlation (DIC). A total of 12 specimens (3 per group) simulating an FP3 maxillary full-arch ISFP supported by four implants were milled from four materials: poly(ether ether ketone) (G1-PEEK), poly(ether ketone ketone) (G2-PEKK), poly(methyl methacrylate) (G3-PMMA), and fiber-reinforced composite (G4-FRC). All specimens were subjected to static loading up to 200 N at the incisors region, corresponding to the anterior unsupported span, and at the occlusal surface of the molars, corresponding to the most distal portion of the cantilever, using a universal testing machine. Full-field vertical displacement and strain distributions (principal tensile, compressive, and von Mises) were acquired through a stereo DIC system and analyzed using a Linear Mixed-Effects Model with Tukey’s HSD post hoc comparisons (α = 0.05). All prostheses withstood the applied load without macroscopic failure. G3-PMMA exhibited the highest vertical displacement, exceeding 1000 µm in the anterior span and 1500 µm in the cantilever region, along with the greatest strain concentrations, particularly at the interproximal embrasures distal to the terminal abutment. G1-PEEK provided the lowest displacement in the anterior span. G4-FRC presented displacements similar to G1-PEEK and G2-PEKK at the distal cantilever, but the lowest tensile strains and the most homogeneous strain dissipation in both loading at the anterior unsupported span and distal cantilever. This indicated that the biomechanical performance of full-arch ISFPs is highly influenced by the polymer used. PEEK, PEKK, and FRC appear as promising alternatives to PMMA for monolithic full-arch rehabilitations.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Coimbra, Centre for Mechanical Engineering, Materials and Processes (CEMMPRE) & Department of Dentistry, Faculty of Medicine, 3000-075 Coimbra, Portugal.
- University of Coimbra, Centre for Mechanical Engineering, Materials and Processes (CEMMPRE) & Department of Mechanical Engineering, Faculty of Sciences and Technology, 3030-788 Coimbra, Portugal.
- Private Practice Implant Centre of the Palm Beaches, West Palm Beach, FL 33401, USA.
- University of Coimbra, Centre for Innovation and Research in Oral Sciences (CIROS) & Department of Dentistry, Faculty of Medicine, 3000-075 Coimbra, Portugal.
ORCID-Profile: Maria Luís Basto, Ana Messias, Maria Augusta Neto, Fernando Guerra, Ana Martins Amaro
Zitieren
APA
Basto, M. L., Messias, A., Neto, M. A., Krauser, J. T., Guerra, F., & Amaro, A. M. (2026). Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study. Polymers, 18(12), 1457. https://doi.org/10.3390/polym18121457
Vancouver
Basto ML, Messias A, Neto MA, Krauser JT, Guerra F, Amaro AM. Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study. Polymers. 2026;18(12):1457. doi:10.3390/polym18121457
BibTeX
@article{Basto2026,
author = {Basto, Maria Luís and Messias, Ana and Neto, Maria Augusta and Krauser, Jack T and Guerra, Fernando and Amaro, Ana Martins},
title = {Biomechanical Evaluation of Cantilevered Full-Arch Implant-Supported Polymer-Based Hybrid Prostheses: A Digital Image Correlation Study},
journal = {Polymers},
year = {2026},
volume = {18},
number = {12},
pages = {1457},
doi = {10.3390/polym18121457}
}
EUVIMED