EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

In-Vivo Dosimetry for Ultra-High Dose Rate (UHDR) Electron Beam FLASH Radiotherapy Using an Organic (Plastic), an Organic–Inorganic Hybrid and an Inorganic Point Scintillator System

Verdi Vanreusel, Hugo Vallet, Jordi Wijnen, Benjamin Côté, Paul Leblans, Paul Sterckx, Dirk Vandenbroucke, Dirk Verellen, Luana de Freitas Nascimento

Photonics · 13.09.2024 · 11 (9) · 865

DOI: 10.3390/photonics11090865 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Photonics, online erschienen am 13.09.2024. Verfasst von Verdi Vanreusel und 8 weiteren Autor:innen. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Belgian Nuclear Research Centre.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Dosimetry is crucial in radiotherapy to warrant safe and correct treatment. In FLASH radiotherapy, where ultra-high dose rates (UHDRs) are used, the dosimetric demands are more stringent, requiring the development and investigation of new dosemeters. In this study, three prototype fiber-optic dosemeters (FODs)—an inorganic, an organic–inorganic hybrid metal halide, and an organic (plastic) scintillator are optimized and investigated for UHDR electron irradiations. The plastic FOD is developed by Medscint, whereas the others are in-house made. The stem signal is minimized by spectral decomposition for the plastic scintillator, and by band-pass wavelength filters for the inorganic and organic–inorganic hybrid metal halide FOD. All prototypes are tested for the dose rate defining parameters. The optimal band-pass wavelength filters are found to be centered around 500 nm and 425 nm for the inorganic and organic–inorganic hybrid metal halide FODs, respectively. A sampling frequency of 1000 Hz is chosen for the inorganic and organic–inorganic hybrid metal halide FODs. The plastic FOD shows to be the least dose rate dependent with maximum deviations of 3% from the reference for the relevant beam settings. The inorganic and organic–inorganic hybrid metal halide FODs, in contrast, show large deviations of >10% from the reference and require more investigation. The current FOD prototypes are insufficient for application in UHDR electron beams, and require further development and investigation.

Förderung

  • Belgian Nuclear Research Centre

Zitieren

APA

Vanreusel, V., Vallet, H., Wijnen, J., Côté, B., Leblans, P., Sterckx, P., Vandenbroucke, D., Verellen, D., & Nascimento, L. D. F. (2024). In-Vivo Dosimetry for Ultra-High Dose Rate (UHDR) Electron Beam FLASH Radiotherapy Using an Organic (Plastic), an Organic–Inorganic Hybrid and an Inorganic Point Scintillator System. Photonics, 11(9), 865. https://doi.org/10.3390/photonics11090865

Vancouver

Vanreusel V, Vallet H, Wijnen J, Côté B, Leblans P, Sterckx P, et al. In-Vivo Dosimetry for Ultra-High Dose Rate (UHDR) Electron Beam FLASH Radiotherapy Using an Organic (Plastic), an Organic–Inorganic Hybrid and an Inorganic Point Scintillator System. Photonics. 2024;11(9):865. doi:10.3390/photonics11090865

BibTeX
@article{Vanreusel2024,
  author = {Vanreusel, Verdi and Vallet, Hugo and Wijnen, Jordi and Côté, Benjamin and Leblans, Paul and Sterckx, Paul and Vandenbroucke, Dirk and Verellen, Dirk and Nascimento, Luana de Freitas},
  title = {In-Vivo Dosimetry for Ultra-High Dose Rate (UHDR) Electron Beam FLASH Radiotherapy Using an Organic (Plastic), an Organic–Inorganic Hybrid and an Inorganic Point Scintillator System},
  journal = {Photonics},
  year = {2024},
  volume = {11},
  number = {9},
  pages = {865},
  doi = {10.3390/photonics11090865}
}
Kostenlos starten →