System Performance Check (SPC): A Readiness Companion to Multi-Attribute Method (MAM) Analysis in Biopharmaceutical Quality Control
Pharmaceuticals · 06.08.2026 · 19 (8) · 1240
DOI: 10.3390/ph19081240 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pharmaceuticals, online erschienen am 06.08.2026. Verfasst von Jahziel Chase und 6 weiteren Autor:innen. Beteiligte Einrichtung: Center of Bio-Innovation, Northeastern University, 147 South Bedford Street, Burlington, MA 01803, USA. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: The multi-attribute method (MAM) is a powerful LC-MS workflow for site-specific monitoring of critical quality attributes (CQAs) in biopharmaceutical products, but its sensitivity to LC drift, ionization variability, and mass-calibration shifts has slowed adoption in routine quality control. A blank injection confirms that the system is clean; it does not confirm that the system is ready. This work evaluates a system performance check (SPC) workflow, built around a 13-peptide LC/MS reference standard designed against the system-readiness metrics in USP General Chapter 1060, as a readiness assessment performed before MAM sample analysis.
Methods
The workflow was run on a recently installed Agilent 6230C LC/TOF with OpenLab CDS and the integrated MAM for OpenLab CDS, following routine instrument tune and calibration. The 13-peptide mix probes mass accuracy, retention time, peak area and height, in-source fragmentation, methionine oxidation, chromatographic resolution, and sodium and iron adduct formation; it also includes a heavy-to-light peptide pair for a single-point, low-abundance response check and a deamidated peptide pair to monitor a clinically important post-translational modification. Triplicate injections were evaluated against per-target acceptance criteria: mass accuracy (±10 ppm or ±13 ppm for EYK), retention-time %RSD (≤0.5%), and, for a subset of targets, peak-area %RSD (≤10%).
Results
Across the three replicates, all thirteen peptides met every acceptance criterion applied to them. The twelve primary targets met their mass-accuracy and retention-time criteria, and the relative-area monitors (sodium and iron adducts, methionine oxidation, the heavy/light detector check, and the deamidation pair) met theirs. The asparagine deamidation pair was detected and reproducibly measured as a relative-area ratio at the +2 and +3 charge states in every replicate (6.7 to 7.0% at +2 and 7.5 to 7.8% at +3, against a
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center of Bio-Innovation, Northeastern University, 147 South Bedford Street, Burlington, MA 01803, USA.
- Department of Chemistry and Chemical Biology, Northeastern University, 360 Huntington Avenue, Boston, MA 02115, USA.
- Barnett Institute of Chemical and Biological Analysis, Northeastern University, 360 Huntington Avenue, Boston, MA 02115, USA.
- Agilent Technologies, Inc., Santa Clara, CA 95051, USA.
ORCID-Profile: Jahziel Chase, Mike Knierman, Jared Auclair
Zitieren
APA
Chase, J., Sato, M., Slysz, G., Miladi, M., Knierman, M., Barkovich, R., & Auclair, J. (2026). System Performance Check (SPC): A Readiness Companion to Multi-Attribute Method (MAM) Analysis in Biopharmaceutical Quality Control. Pharmaceuticals, 19(8), 1240. https://doi.org/10.3390/ph19081240
Vancouver
Chase J, Sato M, Slysz G, Miladi M, Knierman M, Barkovich R, et al. System Performance Check (SPC): A Readiness Companion to Multi-Attribute Method (MAM) Analysis in Biopharmaceutical Quality Control. Pharmaceuticals. 2026;19(8):1240. doi:10.3390/ph19081240
BibTeX
@article{Chase2026,
author = {Chase, Jahziel and Sato, Melissa and Slysz, Gordon and Miladi, Mahsan and Knierman, Mike and Barkovich, Robert and Auclair, Jared},
title = {System Performance Check (SPC): A Readiness Companion to Multi-Attribute Method (MAM) Analysis in Biopharmaceutical Quality Control},
journal = {Pharmaceuticals},
year = {2026},
volume = {19},
number = {8},
pages = {1240},
doi = {10.3390/ph19081240}
}
EUVIMED