Evaluation of the Stress-Shielding Effect of a PEEK Knee Prosthesis. A Finite Element Study
Osteology · 05.08.2025 · 5 (3) · 24
DOI: 10.3390/osteology5030024 ↗Kurz gefasst
Fachartikel in der Zeitschrift Osteology, online erschienen am 05.08.2025. Verfasst von Mario Ceddia und 3 weiteren Autor:innen. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 71 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The long-term success of total knee arthroplasty (TKA) is often compromised by stress shielding, which can lead to bone resorption and even implant loosening. This study employs finite element analysis (FEA) to compare the stress-shielding effects of a knee prosthesis made from polyether ether ketone (PEEK) with a traditional titanium Ti6Al4V implant on an osteoporotic tibial bone model.
Methods
Stress distribution and the stress-shielding factor (SSF) were evaluated at seven critical points in the proximal tibia under physiological loading conditions.
Results
Results indicate that the PEEK prosthesis yields a more uniform stress transmission, with von Mises stress levels within the optimal 2–3 MPa range for bone maintenance and consistently negative or near-zero SSF values, implying minimal stress shielding. Conversely, titanium implants exhibited significant stress shielding with high positive SSF values across all points. Additionally, stress concentrations on the polyethylene liner were lower and more evenly distributed in the PEEK model, suggesting reduced wear potential.
Conclusions
These findings highlight the biomechanical advantages of PEEK in reducing stress shielding and preserving bone integrity, supporting its potential use to improve implant longevity in TKA. Further experimental and clinical validation are warranted.
Zitieren
APA
Ceddia, M., Morizio, A., Solarino, G., & Trentadue, B. (2025). Evaluation of the Stress-Shielding Effect of a PEEK Knee Prosthesis. A Finite Element Study. Osteology, 5(3), 24. https://doi.org/10.3390/osteology5030024
Vancouver
Ceddia M, Morizio A, Solarino G, Trentadue B. Evaluation of the Stress-Shielding Effect of a PEEK Knee Prosthesis. A Finite Element Study. Osteology. 2025;5(3):24. doi:10.3390/osteology5030024
BibTeX
@article{Ceddia2025,
author = {Ceddia, Mario and Morizio, Arcangelo and Solarino, Giuseppe and Trentadue, Bartolomeo},
title = {Evaluation of the Stress-Shielding Effect of a PEEK Knee Prosthesis. A Finite Element Study},
journal = {Osteology},
year = {2025},
volume = {5},
number = {3},
pages = {24},
doi = {10.3390/osteology5030024}
}