Study of the Agreement of the Apnea–Hypopnea Index Measured Simultaneously by Pressure Transducer via Respiratory Polygraphy and by Thermistor via Polysomnography in Real Time with the Same Individuals

Bich-Ty Tran-Thi, Minh Quach-Thieu, Bao-Ngoc Le-Tran, Duy Nguyen-Duc, Nguyen Tran-Hiep, Thao Nguyen-Thi, Yen-Linh Nguyen-Ngoc, Anh Nguyen-Tuan, Tram Tang-Thi-Thao, Toi Nguyen-Van, Sy Duong-Quy

Journal of Otorhinolaryngology, Hearing and Balance Medicine · 2022

DOI: 10.3390/ohbm3020004 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Otorhinolaryngology, Hearing and Balance Medicine, 2022. Verfasst von Bich-Ty Tran-Thi und 10 weiteren Autor:innen. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Obstructive sleep apnea (OSA) is a common disorder and can lead to many severe complications; however, the majority of patients remain undiagnosed. Although polysomnography (PSG) remains the gold standard of diagnosis, it is usually uncomfortable and costly for patients.

Purpose

The study aims to assess the agreement of the AHI measured by polygraphy (PG) (Philips Alice NightOne) with that of polysomnography (Philips Alice PDx) simultaneously recorded in-lab.

Methods

A total of 11 voluntary participants over 18 years old underwent one night of simultaneous PSG and PG recording in sleep laboratories. Studied parameters (AHI, OAI, CAI, MAI, and minSpO2) were analyzed and reported by the Philips Sleepware G3 software. PSG and PG results were scored by qualified staff.

Results

In terms of AHI, the mean AHI derived from PG was different from that of PSG—7.78 and 2.37 events/h, respectively. A Bland–Altman analysis of the AHI on PSG versus PG showed a mean difference of 5.41; limits of agreement (equal to ±2 standard deviations) were from −6.74 to 17.56. The Bland–Altman analysis showed a slight difference between the two methods, with a mean difference of −0.12 events/h in CAI, 1.35 events/h in OAI, and 0.42 events/h in MAI.

Conclusions

In the population with a low suspicion of OSA, the PG showed a low agreement with the simultaneous PSG in the sleep lab. Therefore, PG should only be used as a screening method. Further studies with sufficient sensors in the expanded populations of OSA are needed.

Zitieren

APA

Tran-Thi, B., Quach-Thieu, M., Le-Tran, B., Nguyen-Duc, D., Tran-Hiep, N., Nguyen-Thi, T., Nguyen-Ngoc, Y., Nguyen-Tuan, A., Tang-Thi-Thao, T., Nguyen-Van, T., & Duong-Quy, S. (2022). Study of the Agreement of the Apnea–Hypopnea Index Measured Simultaneously by Pressure Transducer via Respiratory Polygraphy and by Thermistor via Polysomnography in Real Time with the Same Individuals. Journal of Otorhinolaryngology, Hearing and Balance Medicine. https://doi.org/10.3390/ohbm3020004

Vancouver

Tran-Thi B, Quach-Thieu M, Le-Tran B, Nguyen-Duc D, Tran-Hiep N, Nguyen-Thi T, et al. Study of the Agreement of the Apnea–Hypopnea Index Measured Simultaneously by Pressure Transducer via Respiratory Polygraphy and by Thermistor via Polysomnography in Real Time with the Same Individuals. Journal of Otorhinolaryngology, Hearing and Balance Medicine. 2022. doi:10.3390/ohbm3020004

BibTeX
@article{TranThi2022,
  author = {Tran-Thi, Bich-Ty and Quach-Thieu, Minh and Le-Tran, Bao-Ngoc and Nguyen-Duc, Duy and Tran-Hiep, Nguyen and Nguyen-Thi, Thao and Nguyen-Ngoc, Yen-Linh and Nguyen-Tuan, Anh and Tang-Thi-Thao, Tram and Nguyen-Van, Toi and Duong-Quy, Sy},
  title = {Study of the Agreement of the Apnea–Hypopnea Index Measured Simultaneously by Pressure Transducer via Respiratory Polygraphy and by Thermistor via Polysomnography in Real Time with the Same Individuals},
  journal = {Journal of Otorhinolaryngology, Hearing and Balance Medicine},
  year = {2022},
  doi = {10.3390/ohbm3020004}
}
Kostenlos starten →