Safety Testing of Endovascular Devices In Vitro for Interventional Neuroradiology Under 0.55 T MRI
Neuroimaging · 02.04.2026 · 1 (2) · 7
DOI: 10.3390/neuroimaging1020007 ↗Kurz gefasst
Fachartikel in der Zeitschrift Neuroimaging, online erschienen am 02.04.2026. Verfasst von Adèle L. C. Mackowiak und 10 weiteren Autor:innen. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Lausanne University Hospital (CHUV) Departmental Radiology.
Abstract (Originalsprache)
Background/Objectives: MRI-guided neurovascular interventions could benefit from lower-field systems due to reduced magnetic and radiofrequency hazards. However, safety and practical visibility of commonly used neurointerventional devices at 0.55 T remain insufficiently characterized. We evaluated magnetic field interactions, RF-induced heating, and qualitative device visibility in 11 commercially available and commonly used neurovascular devices on a 0.55 T MRI system.
Methods
Eleven devices, including stent retrievers, guidewires, catheters, and one embolization implant, were tested at 0.55 T. Magnetostatic interactions were quantified using the American Society for Testing and Materials (ASTM)-guided deflection methods for translational force (ASTM-F2052) and a two-string suspension apparatus for torque (adapted from Stoianovici et al.). RF-induced heating was measured in an in vitro perfused cerebral vessel phantom using a 15 min high-specific absorption rate spin echo sequence under static and flow conditions. Qualitative device visibility was assessed using a turbo spin echo (TSE) and balanced steady-state free precession (bSSFP) imaging on each device individually.
Results
Eight of eleven devices passed the translational force test, while three devices (D, E, and G), containing significant ferromagnetic components, failed with deflection angles > 45°. Eight devices passed torque testing, remaining below the critical threshold in all rotation positions; three devices (D, G, and J) failed by exceeding the 54° criterion, including one guidewire and two devices with braided/coiled metallic structures. Under static conditions, RF-induced heating ranged from negligible to 10.4 °C (maximum in device D) and generally decreased under flow; in the flow configuration, temperature rise remained below 2 °C for 6/11 devices. Qualitative imaging performance differed by sequence, with bSSFP enabling improved delineation of device structure (best for devices A, C, and H), whereas devices D, E, F, and J produced extensive signal voids that precluded reliable visualization in both sequences. Overall, three devices satisfied all safety criteria while remaining clearly visible under MRI.
Conclusions
Devices that pass safety thresholds at 0.55 T can serve as candidates for further sequence optimization and preclinical workflow development, enabling the design of low-SAR, device-compatible imaging protocols tailored for neurointerventional workflows. These results provide key safety data supporting the feasibility of MR-guided neurovascular procedures at 0.55 T.
Förderung
- Lausanne University Hospital (CHUV) Departmental Radiology
Zitieren
APA
Mackowiak, A. L. C., Eyre, K., Rapacchi, S., Ledoux, J., Swierdzewska, K., Bartolini, B., Puccinelli, F., Saliou, G., Stuber, M., Roy, C. W., & Hajdu, S. D. (2026). Safety Testing of Endovascular Devices In Vitro for Interventional Neuroradiology Under 0.55 T MRI. Neuroimaging, 1(2), 7. https://doi.org/10.3390/neuroimaging1020007
Vancouver
Mackowiak ALC, Eyre K, Rapacchi S, Ledoux J, Swierdzewska K, Bartolini B, et al. Safety Testing of Endovascular Devices In Vitro for Interventional Neuroradiology Under 0.55 T MRI. Neuroimaging. 2026;1(2):7. doi:10.3390/neuroimaging1020007
BibTeX
@article{Mackowiak2026,
author = {Mackowiak, Adèle L. C. and Eyre, Katerina and Rapacchi, Stanislas and Ledoux, Jean-Baptiste and Swierdzewska, Karolina and Bartolini, Bruno and Puccinelli, Francesco and Saliou, Guillaume and Stuber, Matthias and Roy, Christopher W. and Hajdu, Steven D.},
title = {Safety Testing of Endovascular Devices In Vitro for Interventional Neuroradiology Under 0.55 T MRI},
journal = {Neuroimaging},
year = {2026},
volume = {1},
number = {2},
pages = {7},
doi = {10.3390/neuroimaging1020007}
}
EUVIMED