Preanalytical Quality Evaluation of Low-Volume Citrate Evacuated Blood Collection Tubes—Anticoagulant Solution Volume Accuracy, pH, and Anionic–Cationic Composition
Molecules · 02.05.2026 · 31 (9) · 1516
DOI: 10.3390/molecules31091516 ↗Kurz gefasst
Fachartikel in der Zeitschrift Molecules, online erschienen am 02.05.2026. Verfasst von Nataša Gros und 1 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Chemistry and Chemical Technology, University of Ljubljana, Večna pot 113, SI1000 Ljubljana, Slovenia. Hauptthemen laut MEDLINE: Anticoagulants, Blood Specimen Collection, Citric Acid. Die Arbeit nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Slovenian Research and Innovation Agency.
Abstract (Originalsprache)
Blood collection tubes are widely used medical devices. Inaccurate citrate anticoagulant concentration can influence the results of coagulation tests. The producer’s expertise and responsibility are considered the quality safeguards. However, the tubes undergo changes during their lifecycle, partly due to storage conditions, and the end user or a third party has no comprehensive insight. A methodology is necessary to reveal the tube’s inherent characteristics. We provide insight into the anionic–cationic composition and pH of anticoagulant solutions in commercial tubes using high-performance ion exchange chromatography on a purified water model, making the anticoagulant volume accuracy assessment possible through a direct dye-dilution method. The results revealed differences between the tubes of two producers, Greiner BIO-ONE (A and A(nr)) and BD (C). Tube C has the most accurate anticoagulant amount. Both brands contain buffered citrate. The method of buffer preparation is not a source of interferant for the spectrometric method of the tubes’ quality evaluation. Acetate, formate, chloride, nitrite, sulfate, oxalate, bromide, and nitrate impurities were determined in anticoagulant solutions, all in tube A and some in the others. Tubes C exhibit the highest contamination with cations.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Faculty of Chemistry and Chemical Technology, University of Ljubljana, Večna pot 113, SI1000 Ljubljana, Slovenia.
ORCID-Profile: Nataša Gros
Förderung
- Slovenian Research and Innovation Agency
Zitieren
APA
Gros, N., & Hriberšek, Z. (2026). Preanalytical Quality Evaluation of Low-Volume Citrate Evacuated Blood Collection Tubes—Anticoagulant Solution Volume Accuracy, pH, and Anionic–Cationic Composition. Molecules, 31(9), 1516. https://doi.org/10.3390/molecules31091516
Vancouver
Gros N, Hriberšek Z. Preanalytical Quality Evaluation of Low-Volume Citrate Evacuated Blood Collection Tubes—Anticoagulant Solution Volume Accuracy, pH, and Anionic–Cationic Composition. Molecules. 2026;31(9):1516. doi:10.3390/molecules31091516
BibTeX
@article{Gros2026,
author = {Gros, Nataša and Hriberšek, Zala},
title = {Preanalytical Quality Evaluation of Low-Volume Citrate Evacuated Blood Collection Tubes—Anticoagulant Solution Volume Accuracy, pH, and Anionic–Cationic Composition},
journal = {Molecules},
year = {2026},
volume = {31},
number = {9},
pages = {1516},
doi = {10.3390/molecules31091516}
}
EUVIMED