EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

High-Sensitivity Low-Cost 2.61 GHz DGS Sensor for Non-Invasive Glucose Level Monitoring

Felipe Lucena Souza Medeiros, Alexandre Jean René Serres iD, Georgina Karla de Freitas Serres, Ravania Luciano Martildes iD, Caio Vasconcelos Benigno de Abrantes

Micromachines · 29.04.2026 · 17 (5) · 543

DOI: 10.3390/mi17050543 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Micromachines, online erschienen am 29.04.2026. Verfasst von Felipe Lucena Souza Medeiros und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Federal University of Campina Grande, Campina Grande 58429-900, Brazil. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This work presents a loop-shaped (hairpin) resonator incorporating a defective ground structure (DGS) to enhance sensitivity for monitoring water–glucose solutions. The proposed sensor exhibits two resonant frequencies at 2.61 GHz and 4.07 GHz, with reflection coefficients of −46.60 dB and −23.00 dB, respectively. A set of measurements was conducted to compare the performance of the resonator with and without the DGS under two sample-placement configurations: one with water and water–glucose solutions positioned over the feed lines and metallic resonant elements, and another with the water–glucose solutions placed directly over the ground plane. Among the evaluated cases, the ground-plane configuration proved to be the most advantageous, as it produced no frequency shift while yielding distinct magnitude responses of −41.91 dB, −45.62 dB, −47.74 dB, and −49.69 dB for glucose concentrations of 100, 150, 200, and 250 mg/dL, respectively. Overall, the resonator with the defective ground structure consistently demonstrated higher sensitivity and a more stable response pattern, indicating its strong potential for glucose-level monitoring applications.

Schlagwörter

Stichwörter der Autor:innen

diabetesglucoseglycemic index monitoringlow costmicrowavesnon-invasive sensorresonator

Autor:innen und Institutionen

  • Department of Electrical Engineering, Federal University of Campina Grande, Campina Grande 58429-900, Brazil.

ORCID-Profile: Alexandre Jean René Serres, Ravania Luciano Martildes

Zitieren

APA

Medeiros, F. L. S., Serres, A. J. R., Serres, G. K. F., Martildes, R. L., & Benigno de Abrantes, C. V. (2026). High-Sensitivity Low-Cost 2.61 GHz DGS Sensor for Non-Invasive Glucose Level Monitoring. Micromachines, 17(5), 543. https://doi.org/10.3390/mi17050543

Vancouver

Medeiros FLS, Serres AJR, Serres GKF, Martildes RL, Benigno de Abrantes CV. High-Sensitivity Low-Cost 2.61 GHz DGS Sensor for Non-Invasive Glucose Level Monitoring. Micromachines. 2026;17(5):543. doi:10.3390/mi17050543

BibTeX
@article{Medeiros2026,
  author = {Medeiros, Felipe Lucena Souza and Serres, Alexandre Jean René and Serres, Georgina Karla de Freitas and Martildes, Ravania Luciano and Benigno de Abrantes, Caio Vasconcelos},
  title = {High-Sensitivity Low-Cost 2.61 GHz DGS Sensor for Non-Invasive Glucose Level Monitoring},
  journal = {Micromachines},
  year = {2026},
  volume = {17},
  number = {5},
  pages = {543},
  doi = {10.3390/mi17050543}
}
Kostenlos starten →