Study on Electrical Characteristics and ECG Signal Acquisition Performance of Fabric Electrodes Based on Organizational Structure and Wearing Pressure
Micromachines · 17.07.2025 · 16 (7) · 821
DOI: 10.3390/mi16070821 ↗Kurz gefasst
Fachartikel in der Zeitschrift Micromachines, online erschienen am 17.07.2025. Verfasst von Ming Wang und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Chemical and Printing-Dyeing Engineering, Henan University of Engineering, Zhengzhou 450007, China. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Science and Technology Tackle Key Project of Henan Province, The Science and Technology Guidance Project of China National Textile and Apparel Council, Natural Science Foundation Project of Zhongyuan University of Technology.
Abstract (Originalsprache)
Obtaining stable ECG signals under both static and dynamic conditions, while ensuring comfortable wear, is a prerequisite for fabric-electrode applications. It is necessary to study the wearing pressure of fabric electrodes as well as their organizational structure. In this study, fabric electrodes with different organizational structures (plain weave, twill weave, and satin weave) were prepared using silver-plated nylon conductive yarns as weft yarns and polyester yarns as warp yarns. The electrical characteristics of these structures of fabric electrodes were analyzed under different wearing pressures (2 kPa, 3 kPa, 4 kPa, and 5 kPa), and their effects on the quality of static and dynamic ECG signals acquired from human body were examined. The results showed that the contact impedance of the twill and satin weave structured electrodes with the skin was smaller and more stable than that of the plain weave structured electrodes. Furthermore, when a wearing pressure of 3–4 kPa was applied to the satin-structured electrodes, they not only provided satisfactory comfort but also collected stable static and dynamic ECG signals during daily exercise. These results can provide a reference for the application of fabric electrodes in ECG monitoring devices and an important basis for the design of intelligent ECG clothing.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Chemical and Printing-Dyeing Engineering, Henan University of Engineering, Zhengzhou 450007, China.
- College of Intelligent Textile and Fabric Electronics, Zhongyuan University of Technology, Zhengzhou 450007, China.
ORCID-Profile: Jinli Zhou
Förderung
- Science and Technology Tackle Key Project of Henan Province
- The Science and Technology Guidance Project of China National Textile and Apparel Council
- Natural Science Foundation Project of Zhongyuan University of Technology
- Open Project Fund of Henan Provincial Key Laboratory of Functional Textile Materials
- Graduate Research Innovation Program Project of Zhongyuan University of Technology
- Science and Technology Guidance Project of China National Textile and Apparel Council
Zitieren
APA
Wang, M., Zhou, J., & Zhang, G. (2025). Study on Electrical Characteristics and ECG Signal Acquisition Performance of Fabric Electrodes Based on Organizational Structure and Wearing Pressure. Micromachines, 16(7), 821. https://doi.org/10.3390/mi16070821
Vancouver
Wang M, Zhou J, Zhang G. Study on Electrical Characteristics and ECG Signal Acquisition Performance of Fabric Electrodes Based on Organizational Structure and Wearing Pressure. Micromachines. 2025;16(7):821. doi:10.3390/mi16070821
BibTeX
@article{Wang2025,
author = {Wang, Ming and Zhou, Jinli and Zhang, Ge},
title = {Study on Electrical Characteristics and ECG Signal Acquisition Performance of Fabric Electrodes Based on Organizational Structure and Wearing Pressure},
journal = {Micromachines},
year = {2025},
volume = {16},
number = {7},
pages = {821},
doi = {10.3390/mi16070821}
}
EUVIMED