EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Design and Optimization of Planar Spiral Coils for Powering Implantable Neural Recording Microsystem

Jie Luo, Ruifeng Xue iD, Jiahao Cheong, Xuan Zhang, Lei Yao

Micromachines · 09.06.2023 · 14 (6) · 1221

DOI: 10.3390/mi14061221 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Micromachines, online erschienen am 09.06.2023. Verfasst von Jie Luo und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Microelectronics, Shanghai University, Shanghai 201800, China. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Key R&D Program of China, National Key R&D Program of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents a design and optimization method utilizing inductive coupling coils for wireless power transfer in implantable neural recording microsystems, aiming at maximizing power transfer efficiency, which is essential for reducing externally transmitted power and ensuring biological tissue safety. The modeling of inductive coupling is simplified by combining semi-empirical formulations with theoretical models. By introducing the optimal resonant load transformation, the coil optimization is decoupled from an actual load impedance. The complete design optimization process of the coil parameters is given, which takes the maximum theoretical power transfer efficiency as the objective function. When the actual load changes, only the load transformation network needs to be updated instead of rerunning the entire optimization process. Planar spiral coils are designed to power neural recording implants given the challenges of limited implantable space, stringent low-profile restrictions, high-power transmission requirements and biocompatibility. The modeling calculation, electromagnetic simulation and measurement results are compared. The operating frequency of the designed inductive coupling is 13.56 MHz, the outer diameter of the implanted coil is 10 mm and the working distance between the external coil and the implanted coil is 10 mm. The measured power transfer efficiency is 70%, which is close to the maximum theoretical transfer efficiency of 71.9%, confirming the effectiveness of this method.

Schlagwörter

Stichwörter der Autor:innen

implantableinductive couplingneural recording microsystemplanar spiral coilresonant load transformationwireless power transfer

Autor:innen und Institutionen

  • School of Microelectronics, Shanghai University, Shanghai 201800, China.
  • Shanghai Industrial Technology Research Institute, Shanghai 201899, China.

ORCID-Profile: Ruifeng Xue

Förderung

  • National Key R&D Program of China
  • National Key R&D Program of China

Zitieren

APA

Luo, J., Xue, R., Cheong, J., Zhang, X., & Yao, L. (2023). Design and Optimization of Planar Spiral Coils for Powering Implantable Neural Recording Microsystem. Micromachines, 14(6), 1221. https://doi.org/10.3390/mi14061221

Vancouver

Luo J, Xue R, Cheong J, Zhang X, Yao L. Design and Optimization of Planar Spiral Coils for Powering Implantable Neural Recording Microsystem. Micromachines. 2023;14(6):1221. doi:10.3390/mi14061221

BibTeX
@article{Luo2023,
  author = {Luo, Jie and Xue, Ruifeng and Cheong, Jiahao and Zhang, Xuan and Yao, Lei},
  title = {Design and Optimization of Planar Spiral Coils for Powering Implantable Neural Recording Microsystem},
  journal = {Micromachines},
  year = {2023},
  volume = {14},
  number = {6},
  pages = {1221},
  doi = {10.3390/mi14061221}
}
Kostenlos starten →