Digital Impedance Bridge for Four-Terminal-Pair AC Resistor Calibration up to 20 kHz

Mohamed Ouameur, Renata Vasconcellos, Mohamed Agazar

Metrology · 04.01.2024 · 4 (1) · 1-14

DOI: 10.3390/metrology4010001 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Metrology, online erschienen am 04.01.2024. Verfasst von Mohamed Ouameur und 2 weiteren Autor:innen. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

For this study, a substitution principle-based impedance bridge has been developed to calibrate AC resistors in a four-terminal-pair (4TP) configuration. The calibration is performed in the full complex plane for resistances ranging from 100 mΩ to 400 Ω and frequencies of between 50 Hz and 20 kHz. The automated bridge is based on four resistors associated with two high-impedance stages. The balancing of the bridge is achieved by means of PXI modules. The bridge is automatically balanced via a simplex top-down algorithm. The new bridge is primarily used for the measurement chain of AC standard resistors defined in a 4TP configuration at LNE, which are used for routine customer calibrations. The traceability of LNE’s standard resistors when defined in a 4TP configuration is ensured by a measurement chain from a 1 kΩ reference resistor using the new bridge. The reference resistor was calibrated previously via comparison with a calculable resistor up to 20 kHz. The bridge was validated via comparison with calibration results obtained in 1983 and 2009. For a resistor of 1 Ω at 1 kHz, the uncertainty of the series resistance variation and the phase shift are less than 6 µΩ/Ω (k = 1) and 6 µrad (k = 1), respectively.

Zitieren

APA

Ouameur, M., Vasconcellos, R., & Agazar, M. (2024). Digital Impedance Bridge for Four-Terminal-Pair AC Resistor Calibration up to 20 kHz. Metrology, 4(1), 1-14. https://doi.org/10.3390/metrology4010001

Vancouver

Ouameur M, Vasconcellos R, Agazar M. Digital Impedance Bridge for Four-Terminal-Pair AC Resistor Calibration up to 20 kHz. Metrology. 2024;4(1):1-14. doi:10.3390/metrology4010001

BibTeX
@article{Ouameur2024,
  author = {Ouameur, Mohamed and Vasconcellos, Renata and Agazar, Mohamed},
  title = {Digital Impedance Bridge for Four-Terminal-Pair AC Resistor Calibration up to 20 kHz},
  journal = {Metrology},
  year = {2024},
  volume = {4},
  number = {1},
  pages = {1--14},
  doi = {10.3390/metrology4010001}
}
Kostenlos starten →