Evaluation of Cell Mimics as Potential Quality Controls for Human Leukocyte Immunophenotyping
Medical Sciences · 2025
DOI: 10.3390/medsci13040314 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Sciences, 2025. Verfasst von Louis Waeckel und 8 weiteren Autor:innen. Beteiligte Einrichtung: Laboratoire d'Immunologie, CHU Saint-Etienne, 42055 Saint-Etienne Cedex 2, France. Hauptthemen laut MEDLINE: Immunophenotyping, Leukocytes. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Routine lymphocyte counting requires quality assurance validation using quality controls (QCs) that closely resemble fresh human blood leukocytes. This study aimed to evaluate a novel artificial product designed to mimic leukocyte light scatters and marker expression.
Methods
FlowCytes and TruCytes, “artificial cell mimics” (Slingshot Biosciences, Emeryville, CA, USA), were tested on CE-IVD-certified systems, namely FACSCanto, FACSLyric (BD Biosciences), Navios, and DxFlex (Beckman Coulter), using routine staining, lysing, fixation, and no-wash procedures for T, B, and NK counting.
Results
FlowCytes and TruCytes provided forward and side scatter profiles comparable to human leukocytes on the FACSCanto, FACSLyric, and DxFlex systems but not on Navios despite adapting the process to be slightly different from the manufacturer’s recommendations. TruCytes demonstrated robust immunolabeling of CD3, CD4, CD8, and CD19 on the FACSCanto, FACSLyric, and DxFlex systems, with fluorescence intensities and subset distributions being similar to those usually observed in fresh human blood. However, CD16 and CD56 labeling was inconsistent and depended on the antibody clones used. Regrettably, monocyte and granulocyte mimics lacked expression of CD4, CD16, and CD14. TruCytes also displayed significantly lower concentrations of TBNK lymphocyte subsets compared to healthy human blood.
Conclusions
FlowCytes and TruCytes show promises as internal quality controls for T cell and B cell immunophenotyping, but not NK cells. They are compatible with most CE-IVD cytometers, even when using lysis/fixation/no-wash routine diagnosis procedures. Further multicentric studies are warranted to assess their performance relative to existing products, such as stabilized human blood.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratoire d'Immunologie, CHU Saint-Etienne, 42055 Saint-Etienne Cedex 2, France.
- Laboratoire d'Immunologie, CHU Caen, 14000 Caen, France.
- Institut d'Immunologie Laboratoire de Déficits Immunitaires et Immunothérapies, CHRU de Lille, F-59000 Lille, France.
- Laboratoire d'Immunologie et d'Immunogénétique, CHU Limoges, 87042 Limoges, France.
- Laboratoire d'Hématologie et Immunologie, APHP, Hôpital Bichat, 75018 Paris, France.
- Laboratoire d'Immunologie, CHRU Bretonneau Tours, 37044 Tours, France.
ORCID-Profile: Louis Waeckel, Julie Demaret, Gwladys Bourdenet
Zitieren
APA
Waeckel, L., Le Mauff, B., Trauet, J., Demaret, J., Boumediene, A., Hurtado-Nedelec, M., Ciree, A., Bourdenet, G., & Lambert, C. (2025). Evaluation of Cell Mimics as Potential Quality Controls for Human Leukocyte Immunophenotyping. Medical Sciences. https://doi.org/10.3390/medsci13040314
Vancouver
Waeckel L, Le Mauff B, Trauet J, Demaret J, Boumediene A, Hurtado-Nedelec M, et al. Evaluation of Cell Mimics as Potential Quality Controls for Human Leukocyte Immunophenotyping. Medical Sciences. 2025. doi:10.3390/medsci13040314
BibTeX
@article{Waeckel2025,
author = {Waeckel, Louis and Le Mauff, Brigitte and Trauet, Jacques and Demaret, Julie and Boumediene, Ahmed and Hurtado-Nedelec, Margarita and Ciree, Arnaud and Bourdenet, Gwladys and Lambert, Claude},
title = {Evaluation of Cell Mimics as Potential Quality Controls for Human Leukocyte Immunophenotyping},
journal = {Medical Sciences},
year = {2025},
doi = {10.3390/medsci13040314}
}
EUVIMED