EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Performance Study of Anticoagulants and Animal Blood for Establishment of In Vitro Blood Circulation Loop System

Jeonghwa Kim, Taewon Kim

LabMed · 2025

DOI: 10.3390/labmed2020007 ↗

Kurz gefasst

Fachartikel in der Zeitschrift LabMed, 2025. Verfasst von Jeonghwa Kim und 1 weiteren Autor:innen. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministry of Food and Drug Safety.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In vitro blood circulation loop systems are utilized to assess the hemocompatibility and performance of medical devices that come into contact with blood, in accordance with the international standards ASTM F1830 and ASTM F1841. However, a method for evaluating the specific type of anticoagulant and the blood characteristics of each animal species is necessary to ensure consistent and reliable results.

Methods

Blood was collected from healthy rabbits, pigs, rhesus monkeys, and cynomolgus monkeys to evaluate whole blood preserved in anticoagulants (ACD-A, CPDA-1, and heparin). For each sample, red blood cells were monitored over time, and their morphological characteristics were documented.

Results

The morphological grade of erythrocytes gradually decreased over time. Significant differences were observed based on the type of anticoagulant used in the experiment, and variations were noted among different animal species.

Conclusions

The hemocompatibility of in vitro blood circulation loop systems may vary depending on the animal species. Observing erythrocyte morphology can serve as a control measure to ensure reproducible results.

Förderung

  • Ministry of Food and Drug Safety

Zitieren

APA

Kim, J., & Kim, T. (2025). Performance Study of Anticoagulants and Animal Blood for Establishment of In Vitro Blood Circulation Loop System. LabMed. https://doi.org/10.3390/labmed2020007

Vancouver

Kim J, Kim T. Performance Study of Anticoagulants and Animal Blood for Establishment of In Vitro Blood Circulation Loop System. LabMed. 2025. doi:10.3390/labmed2020007

BibTeX
@article{Kim2025,
  author = {Kim, Jeonghwa and Kim, Taewon},
  title = {Performance Study of Anticoagulants and Animal Blood for Establishment of In Vitro Blood Circulation Loop System},
  journal = {LabMed},
  year = {2025},
  doi = {10.3390/labmed2020007}
}
Kostenlos starten →