Comparative Assessment of Pre-Dialysis Fluid Status Using Clinical, Lung, and IVC Ultrasound Methods with Bioimpedance Spectroscopy as Reference Standard in Hemodialysis Patients
Kidney and Dialysis · 12.11.2026 · 6 (4) · 71
DOI: 10.3390/kidneydial6040071 ↗Kurz gefasst
Fachartikel in der Zeitschrift Kidney and Dialysis, online erschienen am 12.11.2026. Verfasst von Ajith M. Nayak und 8 weiteren Autor:innen. Die Arbeit nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Indian Council of Medical Research.
Abstract (Originalsprache)
Background
Accurate assessment of overhydration in maintenance hemodialysis (MHD) patients remains challenging. The clinical method is widely used but may be unreliable. This study compared the nephrologist’s clinical method, lung ultrasound (LUS), and inferior vena cava (IVC) ultrasound against bioimpedance spectroscopy using a body composition monitor (BCM).
Methods
In this single-center cross-sectional study, 98 hypertensive patients receiving twice-weekly MHD were evaluated within 30 min before the second dialysis session of the week. LUS B-lines, IVC diameter index at expiration (IVCDimax), and IVC collapsibility index (IVCCi) were measured. BCM-defined overhydration (>2.5 L) was the reference standard.
Results
BCM classified 88/98 (89.8%) patients as overhydrated; among these, LUS, IVC ultrasound, and nephrologist’s clinical classification identified 86, 82, and 65 patients, respectively. LUS B-lines showed the strongest correlation with BCM-defined overhydration (ρ = 0.913) and the highest observed AUC (0.985; exploratory cutoff ≥12 B-lines: sensitivity 97.7%, specificity 90.0%). IVCCi showed a strong inverse correlation (r = −0.830; AUC = 0.940), while IVCDimax showed a moderate correlation (ρ = 0.641; AUC = 0.853). Clinically estimated overhydration showed a strong correlation (r = 0.797) but lower discriminative ability (AUC = 0.679) and a mean bias of −0.887 L.
Conclusions
LUS had the highest observed AUC; nephrologist’s clinical classification missed more BCM-defined overhydrated patients than LUS or IVC ultrasound.
Förderung
- Indian Council of Medical Research
Zitieren
APA
Nayak, A. M., Prabhu, A. R., Rao, I. R., Bhojaraja, M. V., Prabhu, S., Naik, B., Rangaswamy, D., Shenoy, S. V., & Nagaraju, S. P. (2026). Comparative Assessment of Pre-Dialysis Fluid Status Using Clinical, Lung, and IVC Ultrasound Methods with Bioimpedance Spectroscopy as Reference Standard in Hemodialysis Patients. Kidney and Dialysis, 6(4), 71. https://doi.org/10.3390/kidneydial6040071
Vancouver
Nayak AM, Prabhu AR, Rao IR, Bhojaraja MV, Prabhu S, Naik B, et al. Comparative Assessment of Pre-Dialysis Fluid Status Using Clinical, Lung, and IVC Ultrasound Methods with Bioimpedance Spectroscopy as Reference Standard in Hemodialysis Patients. Kidney and Dialysis. 2026;6(4):71. doi:10.3390/kidneydial6040071
BibTeX
@article{Nayak2026,
author = {Nayak, Ajith M. and Prabhu, Attur Ravindra and Rao, Indu Ramachandra and Bhojaraja, Mohan V. and Prabhu, Shwetha and Naik, Bharathi and Rangaswamy, Dharshan and Shenoy, Srinivas Vinayak and Nagaraju, Shankar Prasad},
title = {Comparative Assessment of Pre-Dialysis Fluid Status Using Clinical, Lung, and IVC Ultrasound Methods with Bioimpedance Spectroscopy as Reference Standard in Hemodialysis Patients},
journal = {Kidney and Dialysis},
year = {2026},
volume = {6},
number = {4},
pages = {71},
doi = {10.3390/kidneydial6040071}
}
EUVIMED