EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Mechanical Thrombectomy with the Vecta 46 Catheter: A Safety and Outcome Analysis

Hunter Hutchinson, Chloe DeYoung, Danyas Sarathy, Grace Hey, Wiley Gillam, Shawna Amini, Muhammad Abdul Baker Chowdhury, Brandon Lucke-Wold, Zachary Sorrentino, Matthew Koch

Journal of Vascular Diseases · 27.04.2026 · 5 (3) · 20

DOI: 10.3390/jvd5030020 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Vascular Diseases, online erschienen am 27.04.2026. Verfasst von Hunter Hutchinson und 9 weiteren Autor:innen. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The AXS Vecta 46 intermediate catheter (Stryker) features a large inner diameter, enabling effective aspiration, combined with a small outer diameter and soft distal-tip flexibility, which allows for safe tracking through more tortuous and smaller arterial segments to target medium vessel occlusions non-traumatically. The efficacy of the Vecta 46 in the spectrum of large and medium vessel occlusions has not been well elucidated in the literature.

Methods

This retrospective cohort study included patients who underwent MT for acute ischemic stroke at our institution between July 2022 and June 2024. The outcomes of patients treated with Vecta 46 were compared to those of all other catheters used at the institution.

Results

The distribution of aspiration and stent retriever attempts in Vecta 46 procedures versus non-Vecta 46 procedures was significantly different (p = 0.00325). Aspiration was attempted 1.66 ± 0.936 times in the Vecta 46 group and 1.12 ± 0.650 times in the non-Vecta 46 group (p = 0.00135). More mechanical thrombectomies with the Vecta 46 included aspiration of a secondary thrombus (p = 0.0314), despite no difference in the distribution of the primary or secondary occlusion location. There were no statistically significant differences in recanalization success (p = 0.800), recanalization time (p = 0.245), procedure duration (p = 0.580), discharge modified Rankin Score (p = 0.875), or intracranial hemorrhage rate (p = 0.720) between non-Vecta 46 and Vecta 46 procedures.

Conclusions

Vecta 46 has similar safety and functional outcomes compared to other endovascular treatment options despite procedural differences.

Zitieren

APA

Hutchinson, H., DeYoung, C., Sarathy, D., Hey, G., Gillam, W., Amini, S., Chowdhury, M. A. B., Lucke-Wold, B., Sorrentino, Z., & Koch, M. (2026). Mechanical Thrombectomy with the Vecta 46 Catheter: A Safety and Outcome Analysis. Journal of Vascular Diseases, 5(3), 20. https://doi.org/10.3390/jvd5030020

Vancouver

Hutchinson H, DeYoung C, Sarathy D, Hey G, Gillam W, Amini S, et al. Mechanical Thrombectomy with the Vecta 46 Catheter: A Safety and Outcome Analysis. Journal of Vascular Diseases. 2026;5(3):20. doi:10.3390/jvd5030020

BibTeX
@article{Hutchinson2026,
  author = {Hutchinson, Hunter and DeYoung, Chloe and Sarathy, Danyas and Hey, Grace and Gillam, Wiley and Amini, Shawna and Chowdhury, Muhammad Abdul Baker and Lucke-Wold, Brandon and Sorrentino, Zachary and Koch, Matthew},
  title = {Mechanical Thrombectomy with the Vecta 46 Catheter: A Safety and Outcome Analysis},
  journal = {Journal of Vascular Diseases},
  year = {2026},
  volume = {5},
  number = {3},
  pages = {20},
  doi = {10.3390/jvd5030020}
}
Kostenlos starten →