Value of Mini Electrodes for Mapping Myocardial Arrhythmogenic Substrate—The Influence of Tip-to-Tissue Angulation and Irrigation Flow on Signal Quality

Karen Bickel, Thorsten Lewalter, Johannes Fischer, Christine Baumgartner, Petra Hoppmann, Klaus Tiemann, Clemens Jilek

Journal of Vascular Diseases · 2022

DOI: 10.3390/jvd1010002 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Vascular Diseases, 2022. Verfasst von Karen Bickel und 6 weiteren Autor:innen. Die Arbeit nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Boston Scientific.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The use of mini electrodes with a small surface and narrow electrode-to-electrode spacing is believed to lead to a higher electrical resolution. Until now, the effects of tip-to-tissue contact, angulation, and irrigation on signal quality and morphology are unknown.

Methods

The beating heart of an open-chest pig was examined while controlling the angulation and contact between the catheter tip and myocardial tissue, as well as the irrigation of the catheter tip. The mini electrodes were mounted onto commercially available 8 mm non-irrigated and 4 mm irrigated tip catheters. Different electrode interconnections, angulations, contact forces, and irrigation flow were analyzed and compared to signals recorded from conventional electrodes.

Results

A total of 63 electrode samples of 21 defined, stable settings, each lasting 30 s, were analyzed. (1) Tissue contact of mini electrodes was given as soon as the conventional tip electrode showed tissue contact. (2) Angulation of the tip-to-tissue contact showed a trend towards changes in the integral of signals derived from mini electrodes, and no significant changes were seen in signals derived from conventional or mini electrodes. (3) Irrigation flow surrounding the mini electrodes did not influence signals derived from mini electrodes, whereas conventional electrodes showed signals with a longer duration under higher irrigation.

Conclusion

Mini electrodes are robust to contact force and irrigation flow regarding signal quality, whereas signals of conventional electrodes are affected by irrigation flow, leading to substantial changes in signal duration and kurtosis. Signals of mini electrodes are sensitive to local electrical changes because of a high local resolution.

Förderung

  • Boston Scientific

Zitieren

APA

Bickel, K., Lewalter, T., Fischer, J., Baumgartner, C., Hoppmann, P., Tiemann, K., & Jilek, C. (2022). Value of Mini Electrodes for Mapping Myocardial Arrhythmogenic Substrate—The Influence of Tip-to-Tissue Angulation and Irrigation Flow on Signal Quality. Journal of Vascular Diseases. https://doi.org/10.3390/jvd1010002

Vancouver

Bickel K, Lewalter T, Fischer J, Baumgartner C, Hoppmann P, Tiemann K, et al. Value of Mini Electrodes for Mapping Myocardial Arrhythmogenic Substrate—The Influence of Tip-to-Tissue Angulation and Irrigation Flow on Signal Quality. Journal of Vascular Diseases. 2022. doi:10.3390/jvd1010002

BibTeX
@article{Bickel2022,
  author = {Bickel, Karen and Lewalter, Thorsten and Fischer, Johannes and Baumgartner, Christine and Hoppmann, Petra and Tiemann, Klaus and Jilek, Clemens},
  title = {Value of Mini Electrodes for Mapping Myocardial Arrhythmogenic Substrate—The Influence of Tip-to-Tissue Angulation and Irrigation Flow on Signal Quality},
  journal = {Journal of Vascular Diseases},
  year = {2022},
  doi = {10.3390/jvd1010002}
}
Kostenlos starten →