Evaluation of Radiation Dose and Image Quality in Clinical Routine Protocols from Three Different CT Scanners

Thawatchai Prabsattroo iD, Jiranthanin Phaorod, Piyaphat Tathuwan, Khanitta Tongluan, Puengjai Punikhom, Tongjit Maharantawong, Waraporn Sudchai

Journal of Imaging · 25.02.2025 · 11 (3) · 70

DOI: 10.3390/jimaging11030070 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Imaging, online erschienen am 25.02.2025. Verfasst von Thawatchai Prabsattroo und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Faculty of Medicine, Khon Kaen University, Khon Kaen 40002, Thailand. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Invitation research grant, Faculty of Medicine, Khon Kaen University, invitation research, Faculty of Medicine, Khon Kaen University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Computed tomography examination plays a vital role in imaging and its use has rapidly increased in radiology diagnosis. This study aimed to assess radiation doses of routine CT protocols of the brain, chest, and abdomen in three different CT scanners, together with a qualitative image quality assessment.

Methods

A picture archiving and communication system (PACS) and Radimetrics software version 3.4.2 retrospectively collected patients’ radiation doses. Radiation doses were recorded as the CTDIvol, dose length product, and effective dose. CT images were acquired using the Catphan700 phantom to evaluate image quality.

Results

The findings revealed that median values for the CTDIvol and DLP across the brain, chest, and abdomen protocols were lower than the national and international DRLs. Effective doses for brain, chest, and abdomen protocols were also below the median value of R. Smith-Bindman. Neusoft achieved higher spatial frequencies in brain protocols, while Siemens outperformed others in chest protocols. Neusoft consistently exhibited superior high-contrast resolution. Siemens and Neusoft outperformed low-contrast detectability, while Siemens also outperformed the contrast-to-noise ratio. In addition, Siemens had the lowest image noise in brain protocols and high uniformity in chest and abdomen protocols. Neusoft showed the lowest noise in chest and abdomen protocols and high uniformity in the brain protocol. The noise power spectrum revealed that Philips had the highest noise magnitude with different noise textures across protocols and scanners.

Conclusions

This study provides a comprehensive evaluation of radiation doses and image quality for three different CT scanners using standard clinical protocols. Almost all CT protocols exhibited radiation doses below the DRLs and demonstrated varying image qualities across each protocol and scanner. Selecting the right CT scanner for each protocol is essential to ensure that the CT images exhibit the best quality among a wide range of CT machines. The MTF, HCR, LCD, CNR, NPS, noise, and uniformity are suitable parameters for evaluating and monitoring image quality.

Schlagwörter

Stichwörter der Autor:innen

high-contrast resolutionimage qualitylow-contrast detectabilityradiation dose

Autor:innen und Institutionen

  • Department of Radiology, Faculty of Medicine, Khon Kaen University, Khon Kaen 40002, Thailand.
  • Nuclear Technology Service Center, Thailand Institute of Nuclear Technology, Nakorn Nayok 26120, Thailand.

ORCID-Profile: Thawatchai Prabsattroo

Förderung

  • Invitation research grant, Faculty of Medicine, Khon Kaen University
  • invitation research, Faculty of Medicine, Khon Kaen University

Zitieren

APA

Prabsattroo, T., Phaorod, J., Tathuwan, P., Tongluan, K., Punikhom, P., Maharantawong, T., & Sudchai, W. (2025). Evaluation of Radiation Dose and Image Quality in Clinical Routine Protocols from Three Different CT Scanners. Journal of Imaging, 11(3), 70. https://doi.org/10.3390/jimaging11030070

Vancouver

Prabsattroo T, Phaorod J, Tathuwan P, Tongluan K, Punikhom P, Maharantawong T, et al. Evaluation of Radiation Dose and Image Quality in Clinical Routine Protocols from Three Different CT Scanners. Journal of Imaging. 2025;11(3):70. doi:10.3390/jimaging11030070

BibTeX
@article{Prabsattroo2025,
  author = {Prabsattroo, Thawatchai and Phaorod, Jiranthanin and Tathuwan, Piyaphat and Tongluan, Khanitta and Punikhom, Puengjai and Maharantawong, Tongjit and Sudchai, Waraporn},
  title = {Evaluation of Radiation Dose and Image Quality in Clinical Routine Protocols from Three Different CT Scanners},
  journal = {Journal of Imaging},
  year = {2025},
  volume = {11},
  number = {3},
  pages = {70},
  doi = {10.3390/jimaging11030070}
}
Kostenlos starten →