EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Optimizing Post-Processing Parameters of 3D-Printed Resin for Surgical Guides

Maria Gabriela Packaeser, Alexander Christiaan Santana, Amanda Maria de Oliveira Dal Piva, Cornelis Johannes Kleverlaan, João Paulo Mendes Tribst

Journal of Composites Science · 10.10.2025 · 9 (10) · 553

DOI: 10.3390/jcs9100553 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Composites Science, online erschienen am 10.10.2025. Verfasst von Maria Gabriela Packaeser und 4 weiteren Autor:innen. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Amsterdam.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluated post-processing protocols for 3D-printed implant surgical guides, aiming to determine the ideal timing after printing and post-curing durations that do not compromise residual monomer release and leachable components or mechanical properties. Specimens made of a surgical guide resin were 3D-printed (Formlabs Form 2) into bars (14 × 1 × 1 mm; n = 10) and square-shaped samples (10 × 10 × 1 mm; n = 1). They were grouped based on the time elapsed after printing (immediate, 24 h, and 72 h) and underwent washing in 99% isopropyl alcohol. Post-curing was performed for 5, 10, 20, or 30 min using a UV-light curing unit (NextDent LC-3DPrint Box). Residual monomer and components levels were assessed through solvent dissolution tests (n = 5), while mechanical properties were evaluated via flexural strength (n = 10) and hardness (n = 10). Statistical analysis with one-way ANOVA and Tukey’s post hoc test showed no significant differences in flexural strength across curing times or storage periods (p > 0.05), with values ranging from 42.93 MPa to 59.43 MPa. Monomers and leachable components were significantly higher immediately after printing (0.84 ± 0.36 mm3) compared to other groups (p < 0.05). For Vickers hardness, a 10 min curing protocol produced values comparable to longer durations (20.26 HV at 20 min/24 h), while the lowest hardness was 14.59 HV in the 5 min groups (p < 0.001). These findings suggest that delaying post-processing up to 72 h and reducing curing time to 10 min do not compromise mechanical properties, released monomers, and leachable components.

Förderung

  • University of Amsterdam

Zitieren

APA

Packaeser, M. G., Santana, A. C., Piva, A. M. D. O. D., Kleverlaan, C. J., & Tribst, J. P. M. (2025). Optimizing Post-Processing Parameters of 3D-Printed Resin for Surgical Guides. Journal of Composites Science, 9(10), 553. https://doi.org/10.3390/jcs9100553

Vancouver

Packaeser MG, Santana AC, Piva AMDOD, Kleverlaan CJ, Tribst JPM. Optimizing Post-Processing Parameters of 3D-Printed Resin for Surgical Guides. Journal of Composites Science. 2025;9(10):553. doi:10.3390/jcs9100553

BibTeX
@article{Packaeser2025,
  author = {Packaeser, Maria Gabriela and Santana, Alexander Christiaan and Piva, Amanda Maria de Oliveira Dal and Kleverlaan, Cornelis Johannes and Tribst, João Paulo Mendes},
  title = {Optimizing Post-Processing Parameters of 3D-Printed Resin for Surgical Guides},
  journal = {Journal of Composites Science},
  year = {2025},
  volume = {9},
  number = {10},
  pages = {553},
  doi = {10.3390/jcs9100553}
}
Kostenlos starten →