One Catheter, Two Chambers: Feasibility and Acute Procedural Outcomes of Dual-Chamber Leadless Pacemaker Implantation with a Single Delivery Catheter
Journal of Clinical Medicine · 06.11.2026 · 15 (21) · 8664
DOI: 10.3390/jcm15218664 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 06.11.2026. Verfasst von Marcin Kowalski und 7 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Institute, Northwell Health, Staten Island University Hospital, New Hyde Park, NY 11040, USA. Die Arbeit nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Dual-chamber leadless pacing with the Aveir DR system restores atrioventricular synchrony without transvenous leads. The labeled implant workflow calls for a separate delivery catheter for each device. The second catheter adds cost and adds steps to the procedure, and whether it is actually needed has not been examined in a large series.
Objective
To describe the acute procedural feasibility, the rate of delivery-catheter malfunction, and the effect on direct disposable supply cost of implanting both devices of a dual-chamber leadless pacemaker system through one delivery catheter.
Methods
We reviewed 102 consecutive patients who received a dual-chamber leadless pacemaker at a single center between February 2024 and February 2025, implanted by five operators at that center. Twenty earlier dual-chamber implants performed with two catheters, before the single-catheter workflow was adopted, precede the study period and are not part of this cohort. The right ventricular device was placed first in every case. The same catheter was then reloaded with the atrial device and used for the second implant, and a new catheter was opened only if the first one failed. We recorded acute implant success, catheter malfunction, procedure time, fluoroscopy time, complications through 30 days after implantation, and direct disposable supply cost.
Results
Both devices were implanted through the original catheter in 100 of 102 patients (98.0%, 95% CI 93.1 to 99.8%). Two patients (2.0%, 95% CI 0.2 to 6.9%) needed a catheter exchange because of a mechanical failure that arose during ventricular deployment; in both, the ventricular device had been recaptured and redeployed with repeated torque beforehand. No catheter failed during atrial implantation. Mean time from sheath insertion to removal was 69 min, and mean fluoroscopy time was 16.0 ± 6.6 min. Two device dislodgements occurred (2.0%, 95% CI 0.2 to 6.9%), one intraprocedural and one a right atrial device that embolized the day after implant; neither was related to the delivery catheter. There was no perforation, pericardial effusion, or access-site complication, and no other complication occurred in any patient through 30 days after implant. Omitting the second catheter lowered direct disposable supply cost by the cost of the second catheter.
Conclusions
With experienced operators, one delivery catheter completed both implants in 98% of cases and lowered disposable supply cost. The study is single arm aand the number of events is small, so these findings support feasibility, with 30-day freedom from complications of 98.0%, and a supply cost difference.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Institute, Northwell Health, Staten Island University Hospital, New Hyde Park, NY 11040, USA.
- Department of Internal Medicine, Staten Island University Hospital, Northwell Health, NY 10305, USA.
ORCID-Profile: Kaiyu Jia
Zitieren
APA
Kowalski, M., Jia, K., Rimsky, E., Kowalski, S., Saouma, S., Shah, R., Akhrass, P., & Parikh, V. (2026). One Catheter, Two Chambers: Feasibility and Acute Procedural Outcomes of Dual-Chamber Leadless Pacemaker Implantation with a Single Delivery Catheter. Journal of Clinical Medicine, 15(21), 8664. https://doi.org/10.3390/jcm15218664
Vancouver
Kowalski M, Jia K, Rimsky E, Kowalski S, Saouma S, Shah R, et al. One Catheter, Two Chambers: Feasibility and Acute Procedural Outcomes of Dual-Chamber Leadless Pacemaker Implantation with a Single Delivery Catheter. Journal of Clinical Medicine. 2026;15(21):8664. doi:10.3390/jcm15218664
BibTeX
@article{Kowalski2026,
author = {Kowalski, Marcin and Jia, Kaiyu and Rimsky, Elizabeth and Kowalski, Sean and Saouma, Samer and Shah, Rina and Akhrass, Philippe and Parikh, Valay},
title = {One Catheter, Two Chambers: Feasibility and Acute Procedural Outcomes of Dual-Chamber Leadless Pacemaker Implantation with a Single Delivery Catheter},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {21},
pages = {8664},
doi = {10.3390/jcm15218664}
}
EUVIMED