Quantification of Peripheral Blur in Ultra-Widefield Fundus Imaging Following Diffractive Multifocal Intraocular Lens Implantation
Journal of Clinical Medicine · 03.08.2026 · 15 (15) · 6044
DOI: 10.3390/jcm15156044 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 03.08.2026. Verfasst von Seong Min Kim und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Informatics, Keimyung University School of Medicine, Daegu 42601, Republic of Korea. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea.
Abstract (Originalsprache)
Background/Objectives: Diffractive multifocal intraocular lenses (IOLs) are widely used following cataract surgery, but their impact on peripheral retinal image quality in ultra-widefield (UWF) fundus imaging has not been objectively characterized. This study aimed to quantify peripheral image sharpness in UWF fundus photographs and assess wavelength-specific differences between diffractive multifocal and enhanced monofocal IOLs.
Methods
In this retrospective cohort study, 65 eyes from 51 patients implanted with enhanced monofocal (Tecnis Eyhance; n = 32) or diffractive multifocal IOLs (Tecnis Synergy, n = 19; Clareon PanOptix, n = 14) were analyzed. Non-mydriatic UWF fundus photographs (Optos; ~200°) were acquired within 1 month postoperatively. Red (R), green (G), and composite channel images were analyzed using three sharpness metrics—Laplacian variance, Brenner gradient, and Tenengrad gradient—within a predefined peripheral annular region (1.5–3.0 times the fovea–optic disc distance). Generalized estimating equations (GEE) with age as a covariate were applied as the primary analysis to account for inter-eye correlation and age differences between groups.
Results
Brenner and Tenengrad gradient scores were significantly lower in the multifocal group in the R channel (p = 0.037 and p = 0.041, respectively), but not in the G or composite channels. The Laplacian metric revealed no significant intergroup differences in any channel. Subgroup analysis revealed no significant difference between hybrid multifocal and quadrifocal subtypes (p > 0.05). Group-level mean heatmaps confirmed a spatially distributed reduction in gradient magnitude across the annular zone in multifocal eyes.
Conclusions
Diffractive multifocal IOL implantation was associated with a localized reduction in peripheral R channel sharpness, hypothesized to reflect transverse chromatic aberration. Composite channel quality remained comparable between groups, suggesting limited impact on routine clinical imaging. These findings may inform IOL selection and interpretation of peripheral retinal findings in multifocal IOL patients.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Informatics, Keimyung University School of Medicine, Daegu 42601, Republic of Korea.
- Department of Ophthalmology, Keimyung University School of Medicine, Daegu 42601, Republic of Korea.
ORCID-Profile: Seong Min Kim, Seung Pil Bang, Seung-Bo Lee
Förderung
- National Research Foundation of Korea
Zitieren
APA
Kim, S. M., Bang, S. P., & Lee, S. B. (2026). Quantification of Peripheral Blur in Ultra-Widefield Fundus Imaging Following Diffractive Multifocal Intraocular Lens Implantation. Journal of Clinical Medicine, 15(15), 6044. https://doi.org/10.3390/jcm15156044
Vancouver
Kim SM, Bang SP, Lee SB. Quantification of Peripheral Blur in Ultra-Widefield Fundus Imaging Following Diffractive Multifocal Intraocular Lens Implantation. Journal of Clinical Medicine. 2026;15(15):6044. doi:10.3390/jcm15156044
BibTeX
@article{Kim2026,
author = {Kim, Seong Min and Bang, Seung Pil and Lee, Seung-Bo},
title = {Quantification of Peripheral Blur in Ultra-Widefield Fundus Imaging Following Diffractive Multifocal Intraocular Lens Implantation},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {15},
pages = {6044},
doi = {10.3390/jcm15156044}
}
EUVIMED