Methodological Limitations of CBCT-Derived Gray Values in Assessing Radiographic Attenuation Patterns After Peri-Implantitis Surgery: Secondary Analysis of a Prospective Clinical Cohort
Journal of Clinical Medicine · 27.05.2026 · 15 (11) · 4144
DOI: 10.3390/jcm15114144 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 27.05.2026. Verfasst von Katarzyna Wieczorek und 6 weiteren Autor:innen. Beteiligte Einrichtung: Chair and Department of Oral Surgery, Medical University of Lublin, 20-093 Lublin, Poland. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medical University of Lublin.
Abstract (Originalsprache)
Objectives
Cone-beam computed tomography (CBCT) is central to three-dimensional assessment in oral surgery and implant dentistry; however, CBCT-derived gray values expressed as HU-like units are not equivalent to true CT-derived Hounsfield Units (HU). This brief methodological secondary analysis evaluated the reliability and practical limitations of such values in assessing radiographic changes after peri-implantitis surgery.
Methods
The analysis used the imaging protocol and group-level radiological data from a previously published prospective clinical cohort, conducted under the same protocol and ethical approval of the Institutional Ethics Committee of the Medical University of Lublin (KE-0254/248/11/2023; 23 November 2023). The source cohort included 57 patients treated after implant removal for severe peri-implantitis with small-particle dentin (n = 22), Bio-Oss (n = 15), or spontaneous healing without grafting (n = 20). CBCT scans were analyzed in OnDemand3D (version 1.0.11.1007) using manually selected square regions of interest (ROI; 30 × 30 pixels). No external phantom calibration, cross-device normalization, or formal intra-/inter-observer reproducibility assessment was available in the secondary dataset.
Results
The previously reported mean study-site values were 779.62 ± 325.92 gray-value units for small-particle dentin, 910.51 ± 155.03 gray-value units for Bio-Oss, and 206.04 ± 174.21 gray-value units for controls. These findings are presented as protocol-dependent attenuation patterns, not as direct material rankings, bone-density thresholds, or proof of regeneration. Variability remained substantial, with study-site coefficients of variation of 41.8%, 17.0%, and 84.6%, respectively, and high adjacent-site variability. Interpretation was constrained by manual ROI placement, lack of calibration, absence of observer-agreement metrics, unequal follow-up timing, and CBCT sensitivity to scatter, beam hardening, field of view, reconstruction settings, and metal-related artifacts.
Conclusions
CBCT-derived gray values may be useful as relative indicators of local radiographic attenuation change within a standardized protocol, but they should not be interpreted as absolute measures of bone density. Future regenerative oral surgery studies should combine standardized acquisition, explicit ROI methodology, repeated measurements, observer-agreement analysis, and complementary clinical, radiographic, or histological outcomes.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Chair and Department of Oral Surgery, Medical University of Lublin, 20-093 Lublin, Poland.
- Department of Experimental Immunology, Medical University of Lublin, 20-093 Lublin, Poland.
- Department of Oral and Maxillofacial Surgery, The Maurice and Gabriela Goldschleger School of Dental Medicine, Tel Aviv University, Tel Aviv 6997801, Israel.
ORCID-Profile: Katarzyna Wieczorek, Grzegorz Hajduk, Michał Łobacz, Paulina Mertowska, Ewelina Grywalska, Sebastian Mertowski, Daya Masri
Förderung
- Medical University of Lublin
Zitieren
APA
Wieczorek, K., Hajduk, G., Łobacz, M., Mertowska, P., Grywalska, E., Mertowski, S., & Masri, D. (2026). Methodological Limitations of CBCT-Derived Gray Values in Assessing Radiographic Attenuation Patterns After Peri-Implantitis Surgery: Secondary Analysis of a Prospective Clinical Cohort. Journal of Clinical Medicine, 15(11), 4144. https://doi.org/10.3390/jcm15114144
Vancouver
Wieczorek K, Hajduk G, Łobacz M, Mertowska P, Grywalska E, Mertowski S, et al. Methodological Limitations of CBCT-Derived Gray Values in Assessing Radiographic Attenuation Patterns After Peri-Implantitis Surgery: Secondary Analysis of a Prospective Clinical Cohort. Journal of Clinical Medicine. 2026;15(11):4144. doi:10.3390/jcm15114144
BibTeX
@article{Wieczorek2026,
author = {Wieczorek, Katarzyna and Hajduk, Grzegorz and Łobacz, Michał and Mertowska, Paulina and Grywalska, Ewelina and Mertowski, Sebastian and Masri, Daya},
title = {Methodological Limitations of CBCT-Derived Gray Values in Assessing Radiographic Attenuation Patterns After Peri-Implantitis Surgery: Secondary Analysis of a Prospective Clinical Cohort},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {11},
pages = {4144},
doi = {10.3390/jcm15114144}
}
EUVIMED