Übersichtsarbeit in der Zeitschrift Journal of Clinical Medicine, online erschienen am 30.04.2026. Verfasst von Viktor Aleinikov und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Spinal Neurosurgery and Peripheral Nervous System Pathology, National Center for Neurosurgery, Astana 010000, Kazakhstan. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: This research was funded by the Science Committee of the Ministry of Science and Higher Education of the Republic of Kazakhstan, Science Committee of the Ministry of Science and Higher Education of the Republic of Kazakhstan.
Abstract (Originalsprache)
Background
Three-dimensional (3D)-printed vertebral body replacement (VBR) and artificial vertebral body (AVB) implants are increasingly used for anterior column reconstruction after spinal tumor resection. However, the available evidence on complications remains limited, heterogeneous, and methodologically inconsistent. This systematic review aimed to synthesize reported complications, revision patterns, and mechanical outcomes of 3D-printed VBR/AVB implants in spinal oncology and to critically appraise the quality of the available clinical literature.
Methods
This systematic review was conducted in accordance with PRISMA 2020. PubMed/MEDLINE, Embase, and the Cochrane Library were searched from 1 January 1980 to 26 February 2026. Eligible studies included clinical series and cohort studies reporting extractable complication and/or revision data in patients who underwent spinal tumor resection followed by reconstruction with a 3D-printed VBR/AVB implant. Methodological quality was assessed using the Joanna Briggs Institute Critical Appraisal Checklist for Case Series. Due to substantial clinical and methodological heterogeneity, a structured narrative synthesis was performed.
Results
Eleven studies comprising 217 analyzable 3D-printed reconstructions were included. Most were retrospective single-center series and showed marked heterogeneity in tumor histology, spinal level, implant strategy, follow-up duration, and complication definitions. Because adverse-event reporting was inconsistent across studies, no pooled overall complication rate was calculated. Reported perioperative non-mechanical complications included neurological deterioration, cerebrospinal fluid- or dural-related events, wound infection, pleural effusion, pneumonia, and vascular injury. Mechanical implant failure appeared relatively uncommon, although radiographic subsidence was variably defined and inconsistently reported. Implant mismatch and hardware-related problems were infrequent but clinically relevant, particularly with prefabricated or off-the-shelf devices. Revision procedures were most commonly associated with wound complications, clinically significant subsidence, hardware failure, or tumor recurrence. Overall study quality was limited by retrospective designs, small sample sizes, and non-standardized outcome reporting.
Conclusions
Current evidence suggests that 3D-printed VBR/AVB implants are a feasible option with encouraging mechanical performance for spinal reconstruction after tumor resection. Most reported adverse events appear to reflect the complexity of oncologic spine surgery rather than device-specific failure alone. However, the available evidence remains low level and heterogeneous. Larger multicenter comparative studies with standardized outcome definitions and longer follow-up are needed to better define the clinical value and durability of 3D-printed vertebral reconstruction in spinal oncology.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Spinal Neurosurgery and Peripheral Nervous System Pathology, National Center for Neurosurgery, Astana 010000, Kazakhstan.
- Department of Radiology and Radiosurgery, National Center for Neurosurgery, Astana 010000, Kazakhstan.
- National Center for Neurosurgery, Astana 010000, Kazakhstan.
ORCID-Profile: Daryn Borangaliyev, Makar Solodovnikov
Förderung
- This research was funded by the Science Committee of the Ministry of Science and Higher Education of the Republic of Kazakhstan
- Science Committee of the Ministry of Science and Higher Education of the Republic of Kazakhstan
Zitieren
APA
Aleinikov, V., Kerimbayev, T., Borangaliyev, D., Kadirbekov, G., Tuigynov, Z., Abishev, N., Zhamoldin, D. K., Oshayev, M., Kenzhegulov, Y., Urunbayev, Y., Baiturlin, Z., Solodovnikov, M., & Akshulakov, S. (2026). Complications and Revision Patterns After 3D-Printed Vertebral Body Replacement for Spinal Tumors: A Systematic Review and Critical Appraisal. Journal of Clinical Medicine, 15(9), 3447. https://doi.org/10.3390/jcm15093447
Vancouver
Aleinikov V, Kerimbayev T, Borangaliyev D, Kadirbekov G, Tuigynov Z, Abishev N, et al. Complications and Revision Patterns After 3D-Printed Vertebral Body Replacement for Spinal Tumors: A Systematic Review and Critical Appraisal. Journal of Clinical Medicine. 2026;15(9):3447. doi:10.3390/jcm15093447
BibTeX
@article{Aleinikov2026,
author = {Aleinikov, Viktor and Kerimbayev, Talgat and Borangaliyev, Daryn and Kadirbekov, Galymzhan and Tuigynov, Zhandos and Abishev, Nurzhan and Zhamoldin, Daniyar K and Oshayev, Meirzhan and Kenzhegulov, Yergen and Urunbayev, Yermek and Baiturlin, Zhanibek and Solodovnikov, Makar and Akshulakov, Serik},
title = {Complications and Revision Patterns After 3D-Printed Vertebral Body Replacement for Spinal Tumors: A Systematic Review and Critical Appraisal},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {9},
pages = {3447},
doi = {10.3390/jcm15093447}
}
EUVIMED