Mid-Term Clinical Outcomes of the Low-Profile Ankura™ Stent Graft System for Endovascular Aneurysm Repair
Journal of Clinical Medicine · 23.04.2026 · 15 (9) · 3231
DOI: 10.3390/jcm15093231 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 23.04.2026. Verfasst von Fatma Akca Ozsar und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Surgery, Kirikkale High Specialization Hospital, 71100 Kirikkale, Turkey. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
To evaluate the real-world safety and mid-term clinical performance of the Ankura™ AAA Stent Graft System in patients undergoing endovascular aneurysm repair (EVAR). Materials and Methods: This prospective, multicenter PMCF study analyzed 100 patients with abdominal aortic aneurysms (AAAs). Patients were monitored for a mean duration of 2.26 years. Primary endpoints included 30-day major adverse events and 24-month treatment success. Statistical evaluation of risk factors for reintervention was performed using univariate logistic regression.
Results
The study cohort was predominantly male (97%), with a mean age of 72.01 years. Hypertension (90%) and smoking (89%) were the most prevalent comorbidities. Regarding the primary endpoints, the 30-day MAE rate was 2%. During the overall follow-up (mean 2.26 years), the primary patency rate was 97%, demonstrating high structural integrity and sustained patency. However, the overall freedom from reintervention rate was 74%, corresponding to a 26% reintervention requirement and a 27% incidence of endoleak. Reinterventions were almost exclusively driven by these post-procedural morphological complications; specifically, 26 of the 27 patients with endoleaks required a secondary procedure. No preoperative clinical or anatomical parameters were identified as significant independent predictors of reintervention in the univariate analysis (p > 0.05). The overall mortality rate was 12%, with 0% aneurysm-related mortality.
Conclusions
Mid-term success and reintervention after EVAR with the Ankura™ system are primarily driven by postoperative morphological complications, such as endoleaks, rather than baseline patient risk profiles. These findings underscore the critical importance of rigorous, lifelong radiological surveillance regardless of preoperative anatomical challenges.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiovascular Surgery, Kirikkale High Specialization Hospital, 71100 Kirikkale, Turkey.
- Department of Cardiovascular Surgery, Ankara Bilkent City Hospital, 06800 Ankara, Turkey.
- Department of Cardiovascular Surgery, Ankara University, 06100 Ankara, Turkey.
ORCID-Profile: Fatma Akca Ozsar, Bekir Bogachan Akkaya, Mehmet Cahit Saricaoglu, Evren Ozcinar
Zitieren
APA
Akca Ozsar, F., Akkaya, B. B., Saricaoglu, M. C., Buyukcakir, O., Ozcinar, E., Iscan, H. Z., & Yazicioglu, L. (2026). Mid-Term Clinical Outcomes of the Low-Profile Ankura™ Stent Graft System for Endovascular Aneurysm Repair. Journal of Clinical Medicine, 15(9), 3231. https://doi.org/10.3390/jcm15093231
Vancouver
Akca Ozsar F, Akkaya BB, Saricaoglu MC, Buyukcakir O, Ozcinar E, Iscan HZ, et al. Mid-Term Clinical Outcomes of the Low-Profile Ankura™ Stent Graft System for Endovascular Aneurysm Repair. Journal of Clinical Medicine. 2026;15(9):3231. doi:10.3390/jcm15093231
BibTeX
@article{AkcaOzsar2026,
author = {Akca Ozsar, Fatma and Akkaya, Bekir Bogachan and Saricaoglu, Mehmet Cahit and Buyukcakir, Onur and Ozcinar, Evren and Iscan, Hakki Zafer and Yazicioglu, Levent},
title = {Mid-Term Clinical Outcomes of the Low-Profile Ankura™ Stent Graft System for Endovascular Aneurysm Repair},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {9},
pages = {3231},
doi = {10.3390/jcm15093231}
}
EUVIMED