EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Radiation Dose Metrics and Local Diagnostic Reference Levels in Low-Dose Stent-Assisted Coiling of Intracranial Aneurysms

Mariusz Stanisław Sowa iD, Joanna Sowa iD, Kamil Adam Węglarz iD, Maciej Budzanowski

Journal of Clinical Medicine · 08.03.2026 · 15 (5) · 2059

DOI: 10.3390/jcm15052059 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 08.03.2026. Verfasst von Mariusz Stanisław Sowa und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurology and Neurosurgery, Faculty of Medicine, Collegium Medicum, University of Warmia and Mazury in Olsztyn, Aleja Warszawska 30, 10…. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Minister of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: Operator experience, the implementation of low frame rates during both fluoroscopy and digital subtraction angiography (DSA), and the use of modern angiographic systems are essential for maintaining diagnostic image quality while minimizing ionizing radiation exposure during stent-assisted endovascular treatment of intracranial aneurysms. At the study center, a low-dose protocol is employed, using the lowest available fluoroscopy frame rate (3.125 frames per second) and a nominal acquisition rate of 2 frames per second for DSA, three-dimensional (3D) rotational angiography, 2D/3D mapping, and roadmapping.

Methods

A retrospective analysis was performed on 132 stent-assisted procedures conducted at a single tertiary center between 2018 and 2024. For each procedure, data were collected for dose-area product (DAP), reference air kerma (Ka,r), fluoroscopy time (FT), and the total number of DSA frames. Local diagnostic reference levels (DRLs; 75th percentile [P75]) and typical values (50th percentile [P50]) were established and compared with values reported in the literature.

Results

For all patients the P75 values, representing DRLs, were 19.89 Gy·cm2 for DAP, 332 mGy for Ka,r, 25 min 32 s for FT, and 354 DSA frames. The P50 values were 13.71 Gy·cm2 for DAP, 219.5 mGy for Ka,r, 20 min 36 s for FT, and 277 DSA frames.

Conclusions

In this single-center cohort, dose metrics for stent-assisted coil embolization were within the lower range of published values. Cross-study comparisons remain descriptive and require cautious interpretation. The proposed local DRLs may support quality assurance, dose optimization, and patient safety in similar clinical settings. Further multicenter and multi-operator studies are necessary to assess transferability and applicability beyond coil-only procedures. Limitations include the retrospective single-center design (single operator) and the lack of a contemporaneous control group and formal image-quality/outcome assessment.

Schlagwörter

Stichwörter der Autor:innen

DRLaneurysm treatmentdiagnostic reference levelsradiation dose optimizationstent-assisted embolization

Autor:innen und Institutionen

  • Department of Neurology and Neurosurgery, Faculty of Medicine, Collegium Medicum, University of Warmia and Mazury in Olsztyn, Aleja Warszawska 30, 10-082 Olsztyn, Poland.
  • Independent Researcher, Aleja Warszawska 30, 10-082 Olsztyn, Poland.
  • Department of Radiation Physics and Dosimetry, Polish Academy of Sciences, 31-342 Kraków, Poland.

ORCID-Profile: Mariusz Stanisław Sowa, Joanna Sowa, Kamil Adam Węglarz

Förderung

  • Minister of Science

Zitieren

APA

Sowa, M. S., Sowa, J., Węglarz, K. A., & Budzanowski, M. (2026). Radiation Dose Metrics and Local Diagnostic Reference Levels in Low-Dose Stent-Assisted Coiling of Intracranial Aneurysms. Journal of Clinical Medicine, 15(5), 2059. https://doi.org/10.3390/jcm15052059

Vancouver

Sowa MS, Sowa J, Węglarz KA, Budzanowski M. Radiation Dose Metrics and Local Diagnostic Reference Levels in Low-Dose Stent-Assisted Coiling of Intracranial Aneurysms. Journal of Clinical Medicine. 2026;15(5):2059. doi:10.3390/jcm15052059

BibTeX
@article{Sowa2026,
  author = {Sowa, Mariusz Stanisław and Sowa, Joanna and Węglarz, Kamil Adam and Budzanowski, Maciej},
  title = {Radiation Dose Metrics and Local Diagnostic Reference Levels in Low-Dose Stent-Assisted Coiling of Intracranial Aneurysms},
  journal = {Journal of Clinical Medicine},
  year = {2026},
  volume = {15},
  number = {5},
  pages = {2059},
  doi = {10.3390/jcm15052059}
}
Kostenlos starten →