Effect of Altered Cervical Thread Pitch on the Primary Stability of Dental Implants
Journal of Clinical Medicine · 21.01.2026 · 15 (2) · 864
DOI: 10.3390/jcm15020864 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 21.01.2026. Verfasst von Lászlo Major und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Surgery, Faculty of Medicine, University of Szeged, 6720 Szeged, Hungary. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 70 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Szeged.
Abstract (Originalsprache)
Background
The macrogeometry and shape of dental implants strongly influence primary stability, which may at times result in excessively high insertion torque. This in vitro study aimed to evaluate whether increasing coronal thread density could reduce insertion torque without compromising primary stability.
Methods
Two conical implants with identical macrogeometry and surface characteristics (Ø 4.2 × 11.5 mm) differed only in the thread pitch of the coronal 3 mm: a modified version (27% more coronal threads; Group 1) and a standard, commercially available version (Group 2). Thirty implants of each design were inserted into high-density (D1; 40 PCF; pounds per cubic foot) and low-density (D3; 20 PCF) polyurethane blocks (n = 120). Insertion torque (IT) and implant stability quotient (ISQ, measured by resonance frequency analysis) were recorded. Group comparisons used the Kruskal–Wallis test, and a generalized linear model (GLM) assessed the independent effects of IT and design on ISQ in D1 bone.
Results
In D1 bone, Group 2 showed higher IT (median 74.0 vs. 63.5 N·cm; p < 0.001) and ISQ (mean 79.1 vs. 77.4; p ≤ 0.030). The GLM identified IT as a negative predictor of ISQ (β = −0.267 per 1 N·cm; p < 0.001), and Group 2 was associated with higher ISQ (+3.90; p < 0.001). In D3 bone, Group 2 again exhibited higher IT (median 37.5 vs. 33.0 N·cm; p < 0.001), while ISQ values were similar between designs (all p > 0.35).
Conclusions
Increasing coronal thread density lowers insertion torque without reducing stability in softer bone and maintains sufficient ISQ for immediate loading in dense bone, making the design advantageous for varied bone qualities.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oral and Maxillofacial Surgery, Faculty of Medicine, University of Szeged, 6720 Szeged, Hungary.
- dicomLAB Dental Ltd., 6720 Szeged, Hungary.
- Department of Operative and Esthetic Dentistry, Faculty of Dentistry, University of Szeged, 6720 Szeged, Hungary.
ORCID-Profile: Mark Adam Antal
Förderung
- University of Szeged
Zitieren
APA
Major, L., Barrak, I., Braunitzer, G., Piffkó, J., & Antal, M. A. (2026). Effect of Altered Cervical Thread Pitch on the Primary Stability of Dental Implants. Journal of Clinical Medicine, 15(2), 864. https://doi.org/10.3390/jcm15020864
Vancouver
Major L, Barrak I, Braunitzer G, Piffkó J, Antal MA. Effect of Altered Cervical Thread Pitch on the Primary Stability of Dental Implants. Journal of Clinical Medicine. 2026;15(2):864. doi:10.3390/jcm15020864
BibTeX
@article{Major2026,
author = {Major, Lászlo and Barrak, Ibrahim and Braunitzer, Gábor and Piffkó, József and Antal, Mark Adam},
title = {Effect of Altered Cervical Thread Pitch on the Primary Stability of Dental Implants},
journal = {Journal of Clinical Medicine},
year = {2026},
volume = {15},
number = {2},
pages = {864},
doi = {10.3390/jcm15020864}
}
EUVIMED