Real-Time Motion Compensation for Dynamic Dental Implant Surgery
Journal of Clinical Medicine · 12.09.2025 · 14 (18) · 6429
DOI: 10.3390/jcm14186429 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 12.09.2025. Verfasst von Daria Pisla und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Maxillofacial Surgery and Radiology, Oral Radiology, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400006 Cluj-Napoca, Romania. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 56 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union-NextGenerationEU and the Romanian Government, under National Recovery and Resilience Plan for Romania, "Iuliu Hatieganu" University of Medicine and Pharmacy Cluj-Napoca, Romania, Department of Maxillofacial Surgery and Radiology, Doctoral Research Program, European Union—NextGenerationEU.
Abstract (Originalsprache)
Background
Accurate and stable instrument positioning is critical in dental implant procedures, particularly in anatomically constrained regions. Conventional navigation systems assume a static patient head, limiting adaptability in dynamic surgical conditions. This study proposes and validates a real-time motion compensation framework that integrates optical motion tracking with a collaborative robot to maintain tool alignment despite patient head movement.
Methods
A six-camera OptiTrack Prime 13 system tracked rigid markers affixed to a 3D-printed human head model. Real-time head pose data were streamed to a Kuka LBR iiwa robot, which guided the implant handpiece to maintain alignment with a predefined target. Motion compensation was achieved through inverse trajectory computation and second-order Butterworth filtering to approximate realistic robotic response. Controlled experiments were performed using the MAiRA Pro M robot to impose precise motion patterns, including pure rotations (±30° at 10–40°/s), pure translations (±50 mm at 5–30 mm/s), and combined sinusoidal motions. Each motion profile was repeated ten times to evaluate intra-trial repeatability and dynamic response.
Results
The system achieved consistent pose tracking errors below 0.2 mm, tool center point (TCP) deviations under 1.5 mm across all motion domains, and an average latency of ~25 ms. Overshoot remained minimal, with effective damping during motion reversal phases. The robot demonstrated stable and repeatable compensation behavior across all experimental conditions.
Conclusions
The proposed framework provides reliable real-time motion compensation for dental implant procedures, maintaining high positional accuracy and stability in the presence of head movement. These results support its potential for enhancing surgical safety and precision in dynamic clinical environments.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Maxillofacial Surgery and Radiology, Oral Radiology, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400006 Cluj-Napoca, Romania.
- Research Centre for Robots Simulation and Testing, Technical University of Cluj-Napoca, 4000114 Cluj-Napoca, Romania.
- Department of Maxillofacial Surgery and Radiology, Maxillofacial Surgery and Implantology, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400429 Cluj-Napoca, Romania.
ORCID-Profile: Daria Pisla, Calin Vaida, Paul Tucan, Cristian Dinu, Doina Pisla
Förderung
- European Union-NextGenerationEU and the Romanian Government, under National Recovery and Resilience Plan for Romania
- "Iuliu Hatieganu" University of Medicine and Pharmacy Cluj-Napoca, Romania, Department of Maxillofacial Surgery and Radiology, Doctoral Research Program
- European Union—NextGenerationEU
- “Iuliu Hatieganu” University of Medicine and Pharmacy Cluj-Napoca, Romania, Department of Maxillofacial Surgery and Radiology, Doctoral Research Program
Zitieren
APA
Pisla, D., Bulbucan, V., Hedeșiu, M., Vaida, C., Cailean, A., Mocan, R., Tucan, P., Dinu, C., Pisla, D., & Team Project Group (2025). Real-Time Motion Compensation for Dynamic Dental Implant Surgery. Journal of Clinical Medicine, 14(18), 6429. https://doi.org/10.3390/jcm14186429
Vancouver
Pisla D, Bulbucan V, Hedeșiu M, Vaida C, Cailean A, Mocan R, et al. Real-Time Motion Compensation for Dynamic Dental Implant Surgery. Journal of Clinical Medicine. 2025;14(18):6429. doi:10.3390/jcm14186429
BibTeX
@article{Pisla2025,
author = {Pisla, Daria and Bulbucan, Vasile and Hedeșiu, Mihaela and Vaida, Calin and Cailean, Andrei and Mocan, Rares and Tucan, Paul and Dinu, Cristian and Pisla, Doina and Team Project Group},
title = {Real-Time Motion Compensation for Dynamic Dental Implant Surgery},
journal = {Journal of Clinical Medicine},
year = {2025},
volume = {14},
number = {18},
pages = {6429},
doi = {10.3390/jcm14186429}
}