EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Assessment of Cardiopulmonary Resuscitation Quality among Healthcare Providers: A Randomized Experimental Study of the Italian Resuscitation Council

Bewertung der Qualität von Herz-Lungen-Wiederbelebung bei Gesundheitsfachpersonen: randomisierte experimentelle Studie des Italian Resuscitation Council

Alberto Cucino, Giovanni Babini iD, Andrea Scapigliati, Giuseppe Ristagno iD

Journal of Clinical Medicine · 15.09.2024 · 13 (18) · 5476

DOI: 10.3390/jcm13185476 ↗

Auf einen Blick

Bei 202 Kongressteilnehmenden verbesserte ein CPR-Feedbackgerät die Qualität der Thoraxkompressionen. Ohne Feedback lag der Anteil korrekter Kompressionen bei etwa 60 %, mit Feedback bei 79 %.

Fragestellung

Welchen Einfluss hat ein CPR-Feedbackgerät auf die Qualität von Thoraxkompressionen bei geschulten Gesundheitsfachpersonen?

Methode

Randomisierte experimentelle Studie mit 202 Kongressteilnehmenden; 2 min Thoraxkompressionen ohne Feedback und 2 min mit Feedback, Reihenfolge randomisiert.

Ergebnisse

Ohne Feedback waren etwa 60 % der Thoraxkompressionen korrekt, mit Feedback 79 %. Im „feedback later“-Arm stiegen korrekte Kompressionen von 51 % auf 86 % und die vollständige Entlastung von 71 % auf 84 %.

Schlussfolgerung der Autor:innen

In dieser Population erwartbar hoch qualifizierter CPR-Anbieter war die Gesamtqualität uneinheitlich und erreichte oft nicht die Leitlinienempfehlungen. Das Feedbackgerät verbesserte die Qualität der Thoraxkompressionen.

Kernaussagen

  • Feedback verbesserte die Qualität der Thoraxkompressionen
  • Ohne Feedback waren etwa 60 % korrekt, mit Feedback 79 %
  • Atemfrequenz blieb in allen Phasen im empfohlenen Bereich

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 15.09.2024. Verfasst von Alberto Cucino und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesia and Intensive Care Medicine 1, Santa Chiara Hospital, Azienda Provinciale per i Servizi Sanitari-APSS Trento, 38122 Trento, …. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The quality of cardiopulmonary resuscitation (CPR) is pivotal in improving the outcome of cardiac arrest. Nevertheless, there is evidence that even professional rescuers may deliver ineffective chest compressions (CCs). We sought to evaluate the impact of a CPR feedback device on the quality of CC performed by a supposedly highly trained and skilled population of attendees to the national annual congress of the Italian Resuscitation Council.

Methods

A total of 202 congress attendees were enrolled to perform 2 min CC without feedback, followed by a 5 min rest and another 2 min interval of CC with feedback. Moreover, participants were randomly assigned to two study groups: “feedback later”, in which the first 2 min CCs were delivered without the feedback, and “feedback first”, in which the first 2 min CCs were aided by feedback. CPR quality has been analyzed in terms of the percentage of CC with adequate depth (CC, %), complete chest release (CR, %), and the CC rate (CC/min).

Results

Approximately 60% of correct CCs were performed without feedback, which significantly increased to 79% with the use of feedback. In the “feedback later” group, the percentage of correctly performed CC and CR significantly increased during the second 2 min interval of CC with feedback (median value 51% vs. 86%, p < 0.0001 and 71% vs. 84%, p = 0.025, respectively). In the “feedback first” group, the percentage of correct CC remained stable during the two CC intervals (71% vs. 80%, p = 0.06), while CR was better without the help of the feedback (70% vs. 93%, p < 0.0001). CC/min was in the recommended range (100–120/min) in all the CC events.

Conclusions

In this population of expected highly skilled CPR providers, the overall quality was inconsistent and, in many cases, did not reach guidelines recommendations. The use of a feedback device significantly improved the quality of CC. When the feedback device was used in the first CC attempt, it had a learning effect that was reflected in maintaining quality during the second CC series.

Schlagwörter

Stichwörter der Autor:innen

CPR feedback deviceCPR qualitycardiac arrestcardiopulmonary resuscitation

Autor:innen und Institutionen

  • Department of Anesthesia and Intensive Care Medicine 1, Santa Chiara Hospital, Azienda Provinciale per i Servizi Sanitari-APSS Trento, 38122 Trento, Italy.
  • Italian Resuscitation Council Steering Committee, 40128 Bologna, Italy.
  • Department of Anesthesiology, Intensive Care and Emergency, Fondazione IRCCS Ca' Granda Ospedale Maggiore Policlinico, 20122 Milan, Italy.
  • Italian Resuscitation Council Scientific Committee, 40128 Bologna, Italy.
  • Institute of Anesthesia and Intensive Care, Catholic University of the Sacred Heart, Fondazione Policlinico Universitario A. Gemelli, IRCCS, 00168 Rome, Italy.
  • Department of Pathophysiology and Transplantation, University of Milan, 20122 Milan, Italy.

ORCID-Profile: Giovanni Babini, Giuseppe Ristagno

Zitieren

APA

Cucino, A., Babini, G., Scapigliati, A., & Ristagno, G. (2024). Assessment of Cardiopulmonary Resuscitation Quality among Healthcare Providers: A Randomized Experimental Study of the Italian Resuscitation Council. Journal of Clinical Medicine, 13(18), 5476. https://doi.org/10.3390/jcm13185476

Vancouver

Cucino A, Babini G, Scapigliati A, Ristagno G. Assessment of Cardiopulmonary Resuscitation Quality among Healthcare Providers: A Randomized Experimental Study of the Italian Resuscitation Council. Journal of Clinical Medicine. 2024;13(18):5476. doi:10.3390/jcm13185476

BibTeX
@article{Cucino2024,
  author = {Cucino, Alberto and Babini, Giovanni and Scapigliati, Andrea and Ristagno, Giuseppe},
  title = {Assessment of Cardiopulmonary Resuscitation Quality among Healthcare Providers: A Randomized Experimental Study of the Italian Resuscitation Council},
  journal = {Journal of Clinical Medicine},
  year = {2024},
  volume = {13},
  number = {18},
  pages = {5476},
  doi = {10.3390/jcm13185476}
}
Kostenlos starten →