Accuracy and Efficacy of Artificial Intelligence-Derived Automatic Measurements of Transthoracic Echocardiography in Routine Clinical Practice
Journal of Clinical Medicine · 24.03.2024 · 13 (7) · 1861
DOI: 10.3390/jcm13071861 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 24.03.2024. Verfasst von Noriko Shiokawa und 6 weiteren Autor:innen. Beteiligte Einrichtung: Ultrasound Center, St. Marianna University Hospital, 2-16-1 Sugao, Miyamae-ku, Kawasaki 216-8511, Japan. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Transthoracic echocardiography (TTE) is the gold standard modality for evaluating cardiac morphology, function, and hemodynamics in clinical practice. While artificial intelligence (AI) is expected to contribute to improved accuracy and is being applied clinically, its impact on daily clinical practice has not been fully evaluated.
Methods
We retrospectively examined 30 consecutive patients who underwent AI-equipped TTE at a single institution. All patients underwent manual and automatic measurements of TTE parameters using the AI-equipped TTE. Measurements were performed by three sonographers with varying experience levels: beginner, intermediate, and expert.
Results
A comparison between the manual and automatic measurements assessed by the experts showed extremely high agreement in the left ventricular (LV) filling velocities (E wave: r = 0.998, A wave: r = 0.996; both p < 0.001). The automated measurements of LV end-diastolic and end-systolic diameters were slightly smaller (−2.41 mm and −1.19 mm) than the manual measurements, although without significant differences, and both methods showing high agreement (r = 0.942 and 0.977, both p < 0.001). However, LV wall thickness showed low agreement between the automated and manual measurements (septum: r = 0.670, posterior: r = 0.561; both p < 0.01), with automated measurements tending to be larger. Regarding interobserver variabilities, statistically significant agreement was observed among the measurements of expert, intermediate, and beginner sonographers for all the measurements. In terms of measurement time, automatic measurement significantly reduced measurement time compared to manual measurement (p < 0.001).
Conclusions
This preliminary study confirms the accuracy and efficacy of AI-equipped TTE in routine clinical practice. A multicenter study with a larger sample size is warranted.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Ultrasound Center, St. Marianna University Hospital, 2-16-1 Sugao, Miyamae-ku, Kawasaki 216-8511, Japan.
- Department of Cardiology, St. Marianna University School of Medicine, 2-16-1 Sugao, Miyamae-ku, Kawasaki 216-8511, Japan.
ORCID-Profile: Masaki Izumo, Yoshihiro Johnny Akashi
Zitieren
APA
Shiokawa, N., Izumo, M., Shimamura, T., Kurosaka, Y., Sato, Y., Okamura, T., & Akashi, Y. J. (2024). Accuracy and Efficacy of Artificial Intelligence-Derived Automatic Measurements of Transthoracic Echocardiography in Routine Clinical Practice. Journal of Clinical Medicine, 13(7), 1861. https://doi.org/10.3390/jcm13071861
Vancouver
Shiokawa N, Izumo M, Shimamura T, Kurosaka Y, Sato Y, Okamura T, et al. Accuracy and Efficacy of Artificial Intelligence-Derived Automatic Measurements of Transthoracic Echocardiography in Routine Clinical Practice. Journal of Clinical Medicine. 2024;13(7):1861. doi:10.3390/jcm13071861
BibTeX
@article{Shiokawa2024,
author = {Shiokawa, Noriko and Izumo, Masaki and Shimamura, Toshio and Kurosaka, Yui and Sato, Yukio and Okamura, Takanori and Akashi, Yoshihiro Johnny},
title = {Accuracy and Efficacy of Artificial Intelligence-Derived Automatic Measurements of Transthoracic Echocardiography in Routine Clinical Practice},
journal = {Journal of Clinical Medicine},
year = {2024},
volume = {13},
number = {7},
pages = {1861},
doi = {10.3390/jcm13071861}
}
EUVIMED