Validation of a Wearable Medical Device for Automatic Diagnosis of OSA against Standard PSG
Journal of Clinical Medicine · 19.01.2024 · 13 (2) · 571
DOI: 10.3390/jcm13020571 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 19.01.2024. Verfasst von Jesus Sanchez Gomez und 4 weiteren Autor:innen. Beteiligte Einrichtung: Sleep Unit, Pneumology Department, Virgen Macarena University Hospital, 41009 Seville, Spain. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Acurable Limited.
Abstract (Originalsprache)
Study objective: The objective of this study was to assess the accuracy of automatic diagnosis of obstructive sleep apnea (OSA) with a new, small, acoustic-based, wearable technology (AcuPebble SA100), by comparing it with standard type 1 polysomnography (PSG) diagnosis. Material and methods: This observational, prospective study was carried out in a Spanish hospital sleep apnea center. Consecutive subjects who had been referred to the hospital following primary care suspicion of OSA were recruited and underwent in-laboratory attended PSG, together with the AcuPebble SA100 device simultaneously overnight from January to December 2022.
Results
A total of 80 patients were recruited for the trial. The patients had a median Epworth scoring of 10, a mean of 10.4, and a range of 0–24. The mean AHI obtained with PSG plus sleep clinician marking was 23.2, median 14.3 and range 0–108. The study demonstrated a diagnostic accuracy (based on AHI) of 95.24%, sensitivity of 92.86%, specificity of 97.14%, positive predictive value of 96.30%, negative predictive value of 94.44%, positive likelihood ratio of 32.50 and negative likelihood ratio of 0.07.
Conclusions
The AcuPebble SA100 (EU) device has demonstrated an accurate automated diagnosis of OSA in patients undergoing in-clinic sleep testing when compared against the gold-standard reference of in-clinic PSG.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Sleep Unit, Pneumology Department, Virgen Macarena University Hospital, 41009 Seville, Spain.
- Wearable Technologies Lab, EEE Department, Imperial College London, London SW7 2AZ, UK.
ORCID-Profile: Syed Anas Imtiaz
Förderung
- Acurable Limited
Zitieren
APA
Sanchez Gomez, J., Pramono, R. X. A., Imtiaz, S. A., Rodriguez-Villegas, E., & Valido Morales, A. (2024). Validation of a Wearable Medical Device for Automatic Diagnosis of OSA against Standard PSG. Journal of Clinical Medicine, 13(2), 571. https://doi.org/10.3390/jcm13020571
Vancouver
Sanchez Gomez J, Pramono RXA, Imtiaz SA, Rodriguez-Villegas E, Valido Morales A. Validation of a Wearable Medical Device for Automatic Diagnosis of OSA against Standard PSG. Journal of Clinical Medicine. 2024;13(2):571. doi:10.3390/jcm13020571
BibTeX
@article{SanchezGomez2024,
author = {Sanchez Gomez, Jesus and Pramono, Renard Xaviero Adhi and Imtiaz, Syed Anas and Rodriguez-Villegas, Esther and Valido Morales, Agustin},
title = {Validation of a Wearable Medical Device for Automatic Diagnosis of OSA against Standard PSG},
journal = {Journal of Clinical Medicine},
year = {2024},
volume = {13},
number = {2},
pages = {571},
doi = {10.3390/jcm13020571}
}
EUVIMED