Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis

Neha Sharma iD, Jokin Zubizarreta-Oteiza iD, Céline Tourbier iD, Florian M Thieringer iD

Journal of Clinical Medicine · 25.03.2023 · 12 (7) · 2495

DOI: 10.3390/jcm12072495 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 25.03.2023. Verfasst von Neha Sharma und 3 weiteren Autor:innen. Beteiligte Einrichtung: Clinic of Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, 4031 Basel, Switzerland. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Werner Siemens Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (p > 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants.

Schlagwörter

Stichwörter der Autor:innen

accuracyalloplastic implantcomputer-assistedcranioplastymaterial extrusionpatient-specificpolymerprintingsterilizationthree-dimensional

Autor:innen und Institutionen

  • Clinic of Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, 4031 Basel, Switzerland.
  • Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Hegenheimermattweg 167C, 4123 Allschwil, Switzerland.

ORCID-Profile: Neha Sharma, Jokin Zubizarreta-Oteiza, Céline Tourbier, Florian M Thieringer

Förderung

  • Werner Siemens Foundation

Zitieren

APA

Sharma, N., Zubizarreta-Oteiza, J., Tourbier, C., & Thieringer, F. M. (2023). Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis. Journal of Clinical Medicine, 12(7), 2495. https://doi.org/10.3390/jcm12072495

Vancouver

Sharma N, Zubizarreta-Oteiza J, Tourbier C, Thieringer FM. Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis. Journal of Clinical Medicine. 2023;12(7):2495. doi:10.3390/jcm12072495

BibTeX
@article{Sharma2023,
  author = {Sharma, Neha and Zubizarreta-Oteiza, Jokin and Tourbier, Céline and Thieringer, Florian M},
  title = {Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis},
  journal = {Journal of Clinical Medicine},
  year = {2023},
  volume = {12},
  number = {7},
  pages = {2495},
  doi = {10.3390/jcm12072495}
}
Kostenlos starten →