Embolization of Perforated Coronary Artery with a Fragment of Balloon Catheter (Cut Balloon Technique)—Multicenter Study
Journal of Cardiovascular Development and Disease · 14.12.2023 · 10 (12) · 496
DOI: 10.3390/jcdd10120496 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Development and Disease, online erschienen am 14.12.2023. Verfasst von Grzegorz Sobieszek und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, 1st Military Hospital, 20-049 Lublin, Poland. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Iatrogenic distal coronary artery perforation can be a life-threatening complication. While there are different dedicated devices for the embolization of distal perforations, there are scarce data about the embolization using the fragmented balloon catheter, the so-called cut balloon technique (CBT).
Methods
We included consecutive patients with distal coronary perforations treated with CBT in four cardiac centers between 2017 and 2023. Clinical, angiographic and procedural characteristics as well as in-hospital outcomes were recorded.
Results
Twenty-six patients (68% men, mean age: 71 ± 10.6 years) with 25 distal coronary perforations and one septal collateral perforation were included. Eleven patients (42%) had elective percutaneous coronary intervention, while fifteen patients (58%) were treated for acute coronary syndrome. The site of perforation was most frequently distributed in the left anterior descending artery (40%), followed by the circumflex artery (28%) and right coronary artery (24%). The diameter of balloons for CBT ranged from 1.5 to 4.0 mm, with most balloons (76%) being either 2.0 or 2.5 mm in diameter. Most balloons (88%) were previously used for lesion predilatation. The numbers of cut balloons needed to seal the perforation were 1, 2 and ≥3 in 48%, 20% and 32% of cases, respectively. The in-hospital prognosis was favorable, with cardiac tamponade requiring pericardiocentesis in only four (16%) patients. Neither emergency surgery nor cardiac death occurred.
Conclusions
CBT is a safe, efficient and easy-to-implement technique for the embolization of coronary perforations. Most distal coronary perforations can be sealed with one or two fragments of cut balloons, obviating the need for additional devices.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, 1st Military Hospital, 20-049 Lublin, Poland.
- Department of Cardiology, Specialistic Hospital, 26-610 Radom, Poland.
- Department of Cardiology, Specialistic Hospital, 22-100 Chełm, Poland.
- Cardiology Division, Santissima Annunziata Hospital, 12038 Savigliano, Italy.
- Department of Interventional Cardiology and Angiology, National Institute of Cardiology, 04-628 Warsaw, Poland.
ORCID-Profile: Grzegorz Sobieszek, Bartosz Zięba, Wojciech Dworzański, Umberto Barbero
Zitieren
APA
Sobieszek, G., Zięba, B., Dworzański, W., Celiński, R., Barbero, U., & Opolski, M. P. (2023). Embolization of Perforated Coronary Artery with a Fragment of Balloon Catheter (Cut Balloon Technique)—Multicenter Study. Journal of Cardiovascular Development and Disease, 10(12), 496. https://doi.org/10.3390/jcdd10120496
Vancouver
Sobieszek G, Zięba B, Dworzański W, Celiński R, Barbero U, Opolski MP. Embolization of Perforated Coronary Artery with a Fragment of Balloon Catheter (Cut Balloon Technique)—Multicenter Study. Journal of Cardiovascular Development and Disease. 2023;10(12):496. doi:10.3390/jcdd10120496
BibTeX
@article{Sobieszek2023,
author = {Sobieszek, Grzegorz and Zięba, Bartosz and Dworzański, Wojciech and Celiński, Rafał and Barbero, Umberto and Opolski, Maksymilian P},
title = {Embolization of Perforated Coronary Artery with a Fragment of Balloon Catheter (Cut Balloon Technique)—Multicenter Study},
journal = {Journal of Cardiovascular Development and Disease},
year = {2023},
volume = {10},
number = {12},
pages = {496},
doi = {10.3390/jcdd10120496}
}
EUVIMED