Surface Plasmon Resonance Immunosensor for Direct Detection of Antibodies against SARS-CoV-2 Nucleocapsid Protein
International Journal of Molecular Sciences · 2024
DOI: 10.3390/ijms25168574 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Molecular Sciences, 2024. Verfasst von Viktorija Lisyte und 4 weiteren Autor:innen. Beteiligte Einrichtung: NanoTechnas-Center of Nanotechnology and Materials Science, Institute of Chemistry, Faculty of Chemistry and Geosciences, Vilnius University, Naugard…. Hauptthemen laut MEDLINE: SARS-CoV-2, Surface Plasmon Resonance, Antibodies, Viral, COVID-19. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research Council of Lithuania and EU, European Regional Development Fund.
Abstract (Originalsprache)
The strong immunogenicity of the SARS-CoV-2 nucleocapsid protein is widely recognized, and the detection of specific antibodies is critical for COVID-19 diagnostics in patients. This research proposed direct, label-free, and sensitive detection of antibodies against the SARS-CoV-2 nucleocapsid protein (anti-SCoV2-rN). Recombinant SARS-CoV-2 nucleocapsid protein (SCoV2-rN) was immobilized by carbodiimide chemistry on an SPR sensor chip coated with a self-assembled monolayer of 11-mercaptoundecanoic acid. When immobilized under optimal conditions, a SCoV2-rN surface mass concentration of 3.61 ± 0.52 ng/mm2 was achieved, maximizing the effectiveness of the immunosensor for the anti-SCoV2-rN determination. The calculated KD value of 6.49 × 10−8 ± 5.3 × 10−9 M confirmed the good affinity of the used monoclonal anti-SCoV2-rN antibodies. The linear range of the developed immunosensor was from 0.5 to 50 nM of anti-SCoV2-rN, where the limit of detection and the limit of quantification values were 0.057 and 0.19 nM, respectively. The immunosensor exhibited good reproducibility and specificity. In addition, the developed immunosensor is suitable for multiple anti-SCoV2-rN antibody detections.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- NanoTechnas-Center of Nanotechnology and Materials Science, Institute of Chemistry, Faculty of Chemistry and Geosciences, Vilnius University, Naugarduko Str. 24, LT-03225 Vilnius, Lithuania.
ORCID-Profile: Viktorija Lisyte, Asta Kausaite-Minkstimiene, Benediktas Brasiunas, Anton Popov, Almira Ramanaviciene
Förderung
- Research Council of Lithuania and EU
- European Regional Development Fund
Zitieren
APA
Lisyte, V., Kausaite-Minkstimiene, A., Brasiunas, B., Popov, A., & Ramanaviciene, A. (2024). Surface Plasmon Resonance Immunosensor for Direct Detection of Antibodies against SARS-CoV-2 Nucleocapsid Protein. International Journal of Molecular Sciences. https://doi.org/10.3390/ijms25168574
Vancouver
Lisyte V, Kausaite-Minkstimiene A, Brasiunas B, Popov A, Ramanaviciene A. Surface Plasmon Resonance Immunosensor for Direct Detection of Antibodies against SARS-CoV-2 Nucleocapsid Protein. International Journal of Molecular Sciences. 2024. doi:10.3390/ijms25168574
BibTeX
@article{Lisyte2024,
author = {Lisyte, Viktorija and Kausaite-Minkstimiene, Asta and Brasiunas, Benediktas and Popov, Anton and Ramanaviciene, Almira},
title = {Surface Plasmon Resonance Immunosensor for Direct Detection of Antibodies against SARS-CoV-2 Nucleocapsid Protein},
journal = {International Journal of Molecular Sciences},
year = {2024},
doi = {10.3390/ijms25168574}
}
EUVIMED