Prototyping and Feasibility Evaluation of a Floor-Embedded Wireless Power Transfer System up to 1.5 kW for Operating Theaters

Akira Tanaka, Kitaro Yoshimitsu, Jun Okamoto, Ken Masamune

International Journal of Medical Devices · 2026

DOI: 10.3390/ijmd1020008 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Medical Devices, 2026. Verfasst von Akira Tanaka und 3 weiteren Autor:innen. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

With the increasing number of cables connected to medical equipment in the healthcare field, the number of medical accidents and incidents (A/Is) is also on the rise. Although wireless power transfer (WPT) for implanted devices has been partially implemented, no studies have reported WPT systems designed to operate along the floor of an operating room. Therefore, in this study, we propose a WPT system for medical equipment in operating rooms, aiming at practical implementation.

Methods

Based on a legal review, on-site observations in operating rooms, and interviews, we defined the system specifications and developed a prototype. We then conducted a system verification of the prototype and performed experiments to supply power to various pieces of medical equipment.

Results

Prototype testing confirmed that the system can deliver up to 1.5 kW (AC 100 V, 15 A); however, due to thermal protection, WPT operation stopped after approximately 30 min at 1.5 kW. Stable operation was demonstrated at 1.2 kW continuously for 8 h from floor-embedded WPT spots. It was also demonstrated that protective earthing can be implemented on the receiving side; nevertheless, the simplified resistance between earth terminals was 0.3 Ω, indicating the need for improvement to satisfy the ≤0.2 Ω requirement for medical locations. With reinforced EMC measures, power could be supplied to noise-sensitive ultrasound imaging equipment without image artifacts.

Conclusions

This study demonstrates the feasibility and remaining challenges of a floor-embedded WPT concept for operating theaters. Key future tasks for practical implementation include achieving continuous 1.5 kW operation with thermal management, meeting protective earthing impedance requirements, and verifying compliance with human exposure guidelines (e.g., ICNIRP).

Zitieren

APA

Tanaka, A., Yoshimitsu, K., Okamoto, J., & Masamune, K. (2026). Prototyping and Feasibility Evaluation of a Floor-Embedded Wireless Power Transfer System up to 1.5 kW for Operating Theaters. International Journal of Medical Devices. https://doi.org/10.3390/ijmd1020008

Vancouver

Tanaka A, Yoshimitsu K, Okamoto J, Masamune K. Prototyping and Feasibility Evaluation of a Floor-Embedded Wireless Power Transfer System up to 1.5 kW for Operating Theaters. International Journal of Medical Devices. 2026. doi:10.3390/ijmd1020008

BibTeX
@article{Tanaka2026,
  author = {Tanaka, Akira and Yoshimitsu, Kitaro and Okamoto, Jun and Masamune, Ken},
  title = {Prototyping and Feasibility Evaluation of a Floor-Embedded Wireless Power Transfer System up to 1.5 kW for Operating Theaters},
  journal = {International Journal of Medical Devices},
  year = {2026},
  doi = {10.3390/ijmd1020008}
}
Kostenlos starten →