High Impedance Active Probe for High Voltages
Hardware · 2024
DOI: 10.3390/hardware2040013 ↗Kurz gefasst
Fachartikel in der Zeitschrift Hardware, 2024. Verfasst von Marc-Aurèle Boillat und 1 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 11 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The probe was designed for the measurement of DC voltages of up to 30 kV from high impedance sources. It is based on a resistive divider with a total resistance of 200 GΩ and a step-down factor of 10’000. In order to allow the measurement of the stepped down voltage with a conventional multimeter without loading, the signal was buffered with an operational amplifier. The device was calibrated against a commercial probe using a low impedance high voltage source. A linear relationship was obtained for a high impedance resistive ladder for voltages between 3 and 30 kV, with a coefficient of determination (R2) of 0.9999. The low-cost device (ca. US $200) fills an application niche not addressed by commercial products.
Zitieren
APA
Boillat, M., & Hauser, P. C. (2024). High Impedance Active Probe for High Voltages. Hardware. https://doi.org/10.3390/hardware2040013
Vancouver
Boillat M, Hauser PC. High Impedance Active Probe for High Voltages. Hardware. 2024. doi:10.3390/hardware2040013
BibTeX
@article{Boillat2024,
author = {Boillat, Marc-Aurèle and Hauser, Peter C.},
title = {High Impedance Active Probe for High Voltages},
journal = {Hardware},
year = {2024},
doi = {10.3390/hardware2040013}
}