Thermal Behavior of Dental Composites During Photopolymerization: Effect of Material Type, Increment Thickness, and Light Intensity

Laura Fontoura, Rim Bourgi, Carlos Enrique Cuevas Suárez, Naji Kharouf, Mohammed Al Hasani, Matías Junge Hess, Abelardo Baez Rosales, Celso Afonso Klein Junior

Eng · 15.05.2026 · 7 (5) · 241

DOI: 10.3390/eng7050241 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Eng, online erschienen am 15.05.2026. Verfasst von Laura Fontoura und 7 weiteren Autor:innen. Die Arbeit nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Heat generated during photopolymerization of resin-based composites from both the exothermic reaction of the material and the irradiance of light-curing units poses a risk to pulp vitality, especially in deep restorations. This study aimed to evaluate temperature variation (ΔT) during the photopolymerization of different resin composites, considering material type, shade, increment thickness, and light-curing unit output. An in vitro experimental study with a factorial design was conducted. Specimens were prepared using 2.0 mm and 4.0 mm increments from conventional (nanohybrid), bulk-fill, and flowable resin composites in different shades (BW, A1, A3, A4, and XB) and different light-curing unit output (100% and 50% battery charge). ΔT was measured using a type K thermocouple (Omega Engineering, Norwalk, CT, USA) positioned at the center of each increment. Data were analyzed using four-way analysis of variance (ANOVA) (α = 0.05). All groups demonstrated a statistically significant temperature increase (p < 0.05), with ΔT values ranging from 3.24 °C to 18.18 °C. Composite type significantly influenced ΔT (p < 0.001), with flowable composites showing the highest temperature rise, followed by bulk-fill and conventional composites. Increment thickness also had a significant effect (p = 0.008), with 4.0 mm increments producing greater temperature increases. Shade significantly affected ΔT (p < 0.001), with the XB shade exhibiting the highest values. Additionally, higher light-curing output (100%) resulted in significantly greater temperature increases compared to 50% output (p < 0.001). Photopolymerization temperature rise is influenced by multiple interacting factors. The combination of flowable composites, darker shades, thicker increments, and higher curing output may increase thermal risk. These findings should be considered when optimizing clinical protocols to minimize potential pulpal damage.

Zitieren

APA

Fontoura, L., Bourgi, R., Suárez, C. E. C., Kharouf, N., Hasani, M. A., Hess, M. J., Rosales, A. B., & Junior, C. A. K. (2026). Thermal Behavior of Dental Composites During Photopolymerization: Effect of Material Type, Increment Thickness, and Light Intensity. Eng, 7(5), 241. https://doi.org/10.3390/eng7050241

Vancouver

Fontoura L, Bourgi R, Suárez CEC, Kharouf N, Hasani MA, Hess MJ, et al. Thermal Behavior of Dental Composites During Photopolymerization: Effect of Material Type, Increment Thickness, and Light Intensity. Eng. 2026;7(5):241. doi:10.3390/eng7050241

BibTeX
@article{Fontoura2026,
  author = {Fontoura, Laura and Bourgi, Rim and Suárez, Carlos Enrique Cuevas and Kharouf, Naji and Hasani, Mohammed Al and Hess, Matías Junge and Rosales, Abelardo Baez and Junior, Celso Afonso Klein},
  title = {Thermal Behavior of Dental Composites During Photopolymerization: Effect of Material Type, Increment Thickness, and Light Intensity},
  journal = {Eng},
  year = {2026},
  volume = {7},
  number = {5},
  pages = {241},
  doi = {10.3390/eng7050241}
}
Kostenlos starten →