Global Solar Irradiance Measurement System Based on a Pyranometer, Data Acquisition Board, and LabVIEW Interface

Adrian-Emanuel Magheț, Dumitru Toader, Simona Ilie

Electronics · 11.07.2026 · 15 (14) · 3057

DOI: 10.3390/electronics15143057 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Electronics, online erschienen am 11.07.2026. Verfasst von Adrian-Emanuel Magheț und 2 weiteren Autor:innen. Die Arbeit nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents a global solar irradiance measurement system developed at Politehnica University Timisoara (UPT), comprising a Kipp & Zonen CMP11 thermopile pyranometer (Kipp & Zonen, OTT HydroMet B.V., Delft, The Netherlands), a National Instruments myDAQ board (National Instruments Corporation, Austin, TX, USA), and a LabVIEW 2018 interface (National Instruments Corporation, Austin, TX, USA). The platform provides configurable acquisition, voltage-to-irradiance conversion, temporal averaging, data storage, and MATLAB R2024b-based analysis (The MathWorks, Inc., Natick, MA, USA). A preliminary operational validation was performed against ten-minute global-irradiance data supplied by the National Meteorological Administration of Romania (ANM) at the same site. The dataset comprised five non-consecutive days between December 2025 and May 2026, with 215 paired values. The global metrics were MBE = −0.33 W/m2, MAE = 4.55 W/m2, RMSE = 6.05 W/m2, nRMSE = 1.29%, and MARE = 0.96%. Incident solar energy was 60.357 MJ/m2 for UPT and 60.399 MJ/m2 for ANM, a relative difference of −0.07%. Day-specific Bland–Altman analyses showed 0.00–4.65% of differences outside the corresponding limits of agreement. Regression of the differences against the pairwise mean irradiance identified an irradiance-dependent bias component in the combined dataset. The results demonstrate promising operational agreement under the tested conditions. Longer campaigns and traceable low-level characterization of the acquisition chain remain necessary for extended metrological assessment.

Zitieren

APA

Magheț, A., Toader, D., & Ilie, S. (2026). Global Solar Irradiance Measurement System Based on a Pyranometer, Data Acquisition Board, and LabVIEW Interface. Electronics, 15(14), 3057. https://doi.org/10.3390/electronics15143057

Vancouver

Magheț A, Toader D, Ilie S. Global Solar Irradiance Measurement System Based on a Pyranometer, Data Acquisition Board, and LabVIEW Interface. Electronics. 2026;15(14):3057. doi:10.3390/electronics15143057

BibTeX
@article{Maghe2026,
  author = {Magheț, Adrian-Emanuel and Toader, Dumitru and Ilie, Simona},
  title = {Global Solar Irradiance Measurement System Based on a Pyranometer, Data Acquisition Board, and LabVIEW Interface},
  journal = {Electronics},
  year = {2026},
  volume = {15},
  number = {14},
  pages = {3057},
  doi = {10.3390/electronics15143057}
}