A Source-Traced PET/CT Scanner Performance Database (2004–2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery

Rajeh Assiri, Ibrahim Hadadi, Mohammed Asiri, Abdulaziz Yahya M. Hussain, Hussein Alshaari, Mohammad Sayed, Mansour M. Alqahtani

Diagnostics · 11.11.2026 · 16 (22) · 3725

DOI: 10.3390/diagnostics16223725 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 11.11.2026. Verfasst von Rajeh Assiri und 6 weiteren Autor:innen. Die Arbeit nennt 75 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: King Khalid University, Najran University.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Small-lesion quantification is a key task in diagnostic PET/CT. However, published performance data span two decades and multiple NEMA NU 2 revisions, complicating comparisons. We developed a source-traced database to examine how reconstruction-associated improvements in hot-sphere contrast recovery vary with sphere diameter, providing reference data for phantom-based protocol optimisation.

Methods

Published NEMA and phantom measurements were source-verified, linked to scanner geometry, and annotated with acquisition and reconstruction conditions. Contrast recovery at a 4:1 sphere-to-background activity ratio (151 measurements; 13 scanners) was analysed using mixed-effects meta-regression with a scanner random intercept.

Results

The database included 28 scanner models (2004–2024), 610 measurements and 43 publications. Descriptively, sensitivity was 3.8–10.9 cps/kBq for lutetium-based analogue scanners, 12.1–47.6 for conventional-geometry digital scanners and 176–192 for total-body scanners. Time of flight (TOF) was associated with a 7.9-percentage-point-higher contrast recovery (95% CI 4.8 to 11.1), but the association varied with sphere diameter (p < 0.001) and was significant only at 10–17 mm (11.4, 9.9 and 7.0 points); estimates at 28 and 37 mm rested on four to five scanners and were imprecise. The median between-laboratory difference under matched conditions was 5.7 points. Point-spread-function and regularised-reconstruction effects were not robust.

Conclusions

In phantom measurements, TOF-associated gains were concentrated at 10–17 mm. These data provide reference values for phantom-based protocol optimisation, harmonisation and acceptance testing; they do not measure lesion detectability or diagnostic accuracy, which require patient-based validation.

Förderung

  • King Khalid University
  • Najran University

Zitieren

APA

Assiri, R., Hadadi, I., Asiri, M., Hussain, A. Y. M., Alshaari, H., Sayed, M., & Alqahtani, M. M. (2026). A Source-Traced PET/CT Scanner Performance Database (2004–2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery. Diagnostics, 16(22), 3725. https://doi.org/10.3390/diagnostics16223725

Vancouver

Assiri R, Hadadi I, Asiri M, Hussain AYM, Alshaari H, Sayed M, et al. A Source-Traced PET/CT Scanner Performance Database (2004–2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery. Diagnostics. 2026;16(22):3725. doi:10.3390/diagnostics16223725

BibTeX
@article{Assiri2026,
  author = {Assiri, Rajeh and Hadadi, Ibrahim and Asiri, Mohammed and Hussain, Abdulaziz Yahya M. and Alshaari, Hussein and Sayed, Mohammad and Alqahtani, Mansour M.},
  title = {A Source-Traced PET/CT Scanner Performance Database (2004–2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery},
  journal = {Diagnostics},
  year = {2026},
  volume = {16},
  number = {22},
  pages = {3725},
  doi = {10.3390/diagnostics16223725}
}
Kostenlos starten →