Effective Dose Evaluation of Cerebral Angiography Using Kerma Area Product and Development of Effective Dose Conversion Factors
Diagnostics · 05.08.2026 · 16 (15) · 2467
DOI: 10.3390/diagnostics16152467 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 05.08.2026. Verfasst von Jeongbok Min und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Bio-Health Convergence Major in Medical Convergence Radiologic-Technology, Graduate School of Health Technology, Daegu Health College, …. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: Transfemoral cerebral angiography (TFCA) involves significant radiation exposure because of prolonged fluoroscopy and repeated imaging sequences. Because complex dose simulations are often infeasible in routine clinical practice, this study employed PC-based Monte Carlo methods to estimate organ-absorbed and effective doses. The primary objective was to establish Kerma area product (KAP)-based conversion factors to facilitate efficient clinical dose assessment.
Methods
We analyzed 30 patients who underwent TFCA. Key parameters, including total KAP, tube voltage, and projection angles, were obtained from DICOM Radiation Dose Structured Reports. Monte Carlo simulations were conducted for anteroposterior, lateral, and cranial planes. The conversion factors were then derived by calculating the ratio of the effective dose to the total KAP.
Results
The average KAP was 36.23 ± 6.81 Gy·cm2. Effective doses calculated via Monte Carlo simulation were 1.70 ± 0.35 mSv under International Commission on Radiological Protection (ICRP) 60 and 1.90 ± 0.40 mSv under ICRP 103 standards. The corresponding conversion factors were 0.0468 ± 0.0054 and 0.0525 ± 0.0065 mSv/Gy·cm2 respectively. Notably, the ICRP 103-based factor was 12.2 percent higher than the ICRP 60-based factor, reflecting updated tissue weighting factors in modern dosimetry.
Conclusions
This study provides standardized conversion factors that allow rapid and reliable dose estimation without requiring complex simulations. The findings demonstrated substantial radiation absorption in the head and neck regions, providing critical data for routine dose monitoring. These results emphasize the importance of rigorous radiation protection and safety protocols during interventional procedures.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Bio-Health Convergence Major in Medical Convergence Radiologic-Technology, Graduate School of Health Technology, Daegu Health College, Daegu 41453, Republic of Korea.
- Department of Radiology, Ulsan Joongang Hospital, Ulsan 44667, Republic of Korea.
- Department of Radiological-Technology, Daegu Health College, Daegu 41453, Republic of Korea.
ORCID-Profile: Jungsu Kim
Zitieren
APA
Min, J., & Kim, J. (2026). Effective Dose Evaluation of Cerebral Angiography Using Kerma Area Product and Development of Effective Dose Conversion Factors. Diagnostics, 16(15), 2467. https://doi.org/10.3390/diagnostics16152467
Vancouver
Min J, Kim J. Effective Dose Evaluation of Cerebral Angiography Using Kerma Area Product and Development of Effective Dose Conversion Factors. Diagnostics. 2026;16(15):2467. doi:10.3390/diagnostics16152467
BibTeX
@article{Min2026,
author = {Min, Jeongbok and Kim, Jungsu},
title = {Effective Dose Evaluation of Cerebral Angiography Using Kerma Area Product and Development of Effective Dose Conversion Factors},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {15},
pages = {2467},
doi = {10.3390/diagnostics16152467}
}