Laser-Assisted Self-Monitoring of Blood Glucose: Analytical Performance, Clinical Accuracy, and Usability of the HandyRay-Glu System
Diagnostics · 31.05.2026 · 16 (11) · 1700
DOI: 10.3390/diagnostics16111700 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 31.05.2026. Verfasst von Minsup Lim und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Health and Safety Convergence Science, School of Health Science, Korea University, Seoul 02841, Republic of Korea. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Korea Health Industry Development Institute.
Abstract (Originalsprache)
Background/Objectives: Diabetes mellitus is a major global health burden, and inadequate glycemic control increases the risk of microvascular and macrovascular complications. Self-monitoring of blood glucose (SMBG) is essential for diabetes management, but conventional finger-prick sampling may reduce adherence due to pain and repeated skin injury. This study evaluated the analytical performance, clinical accuracy, and usability of a novel laser-assisted blood glucose monitoring system, HandyRay-Glu.
Methods
A prospective clinical evaluation study was conducted in accordance with ISO 15197:2013. Capillary blood glucose values obtained using the HandyRay-Glu system were compared with reference measurements generated by the cobas c111 analyzer. Analytical performance was assessed by evaluating repeatability, linearity, hematocrit effect, and interference. Clinical performance was assessed according to ISO 15197:2013 system accuracy criteria, and method comparison was performed using Passing–Bablok regression and Bland–Altman analyses. Usability was evaluated using a structured participant questionnaire.
Results
A total of 100 adult participants with diabetes mellitus were included. Overall, 97.8% of results met the ISO 15197:2013 accuracy criteria. Passing–Bablok regression showed strong agreement between HandyRay-Glu and the reference method (y = 1.694 + 0.9859x, r = 0.992). Bland–Altman analysis demonstrated a mean bias of −1.763 mg/dL, with 95% limits of agreement ranging from −29.333 to 25.808 mg/dL. Analytical evaluations showed acceptable repeatability, linearity across the tested measurement range, and no clinically significant interference. More than 97% of participants reported satisfaction with device usability.
Conclusions
The HandyRay-Glu system met the performance requirements of ISO 15197:2013 and demonstrated high analytical accuracy, acceptable agreement with the reference method, and favorable usability. Laser-assisted blood sampling combined with electrochemical glucose measurement may offer a potential alternative to conventional SMBG systems, and its possible role in improving patient acceptance of regular monitoring warrants further investigation.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Health and Safety Convergence Science, School of Health Science, Korea University, Seoul 02841, Republic of Korea.
- Department of Laboratory Medicine, Korea University Guro Hospital, Korea University College of Medicine, Seoul 02842, Republic of Korea.
- Division of Endocrinology and Metabolism, Department of Internal Medicine, Korea University Ansan Hospital, Korea University College of Medicine, Seoul 02842, Republic of Korea.
- Department of Laboratory Medicine, Korea University Ansan Hospital, Korea University College of Medicine, Seoul 02842, Republic of Korea.
ORCID-Profile: Minsup Lim
Förderung
- Korea Health Industry Development Institute
Zitieren
APA
Lim, M., Lee, J., Seo, J. A., & Ko, S. Y. (2026). Laser-Assisted Self-Monitoring of Blood Glucose: Analytical Performance, Clinical Accuracy, and Usability of the HandyRay-Glu System. Diagnostics, 16(11), 1700. https://doi.org/10.3390/diagnostics16111700
Vancouver
Lim M, Lee J, Seo JA, Ko SY. Laser-Assisted Self-Monitoring of Blood Glucose: Analytical Performance, Clinical Accuracy, and Usability of the HandyRay-Glu System. Diagnostics. 2026;16(11):1700. doi:10.3390/diagnostics16111700
BibTeX
@article{Lim2026,
author = {Lim, Minsup and Lee, JunMin and Seo, Ji A and Ko, Sun-Young},
title = {Laser-Assisted Self-Monitoring of Blood Glucose: Analytical Performance, Clinical Accuracy, and Usability of the HandyRay-Glu System},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {11},
pages = {1700},
doi = {10.3390/diagnostics16111700}
}