Performance of a Screening Mammography AI Algorithm Repurposed for Symptomatic Mammography in a Tertiary Outpatient Clinic

Helen Ngo iD, Eric Niller, Eric Schmitz, Elmar Kotter, Marisa Windfuhr-Blum, Claudia Neubauer, Ana-Luisa Palacios, Fabian Bamberg, Jakob Neubauer iD, Jakob Weiss, Caroline Wilpert iD

Diagnostics · 25.03.2026 · 16 (7) · 984

DOI: 10.3390/diagnostics16070984 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostics, online erschienen am 25.03.2026. Verfasst von Helen Ngo und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, Medical Center-University of Freiburg, Faculty of Medicine-University of Freiburg, 79106 Freib…. Die Arbeit nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Lunit.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: The aim of the study was to evaluate the diagnostic accuracy of a commercial artificial intelligence (AI) algorithm originally developed for screening mammography when applied to symptomatic women presenting to a tertiary outpatient clinic.

Methods

This single-center, retrospective diagnostic accuracy study included women who presented with breast symptoms to a tertiary outpatient clinic between January and June 2013 and underwent digital mammography. An AI algorithm cleared by the U.S. Food and Drug Administration (FDA)-cleared AI algorithm was applied to all mammograms and generated continuous malignancy scores ranging from 1 to 100. Mammographic breast density was classified according to the American College of Radiology Breast Imaging Reporting and Data System (BI-RADS) by two experienced radiologists. Histopathology, when available, or otherwise a minimum of 2 years of clinical and imaging follow-up served as the reference standard. Diagnostic performance was assessed using receiver operating characteristic (ROC) analysis with calculation of the area under the curve (AUC) and 95% confidence intervals (CI) derived by patient level bootstrap resampling (n = 2000). Analyses were performed for the overall cohort and stratified by breast density (non-dense [BI-RADS A–B] vs. dense [BI-RADS C–D]).

Results

A total of 78 women (mean age, 55 ± 11 years) were included, of whom 16 had histopathological verification of suspicious lesions with proven breast cancer in 14 patients and 62 were classified based on follow-up alone. In the overall cohort (156 breasts, including 15 breasts with malignancies), the AI algorithm achieved an AUC of 0.96 (95% CI: 0.86–1.00). Performance remained high in non-dense breasts (AUC = 0.96; 95% CI: 0.88–1.00) and dense breasts (AUC = 0.99; 95% CI: 0.93–1.00), with no statistically significant difference observed between density subgroups (DeLong test, p = 0.36), although subgroup comparisons were underpowered. Decision curve analysis suggested a consistent positive net benefit across a wide range of threshold probabilities in both density groups.

Conclusions

In this preliminary, single-center retrospective cohort, a screening-trained AI algorithm showed promising diagnostic accuracy when applied to symptomatic mammograms. These findings require validation in larger, contemporary, multicenter cohorts before clinical implementation.

Schlagwörter

Stichwörter der Autor:innen

artificial intelligencebreast cancerbreast densitycomputer-aided detectiondeep learningdiagnostic imagingmammographysymptomatic

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Radiology, Medical Center-University of Freiburg, Faculty of Medicine-University of Freiburg, 79106 Freiburg, Germany.
  • Lunit, 7F, 374, Gangnam-daero, Gangnam-gu, Seoul 06241, Republic of Korea.

ORCID-Profile: Helen Ngo, Jakob Neubauer, Caroline Wilpert

Förderung

  • Lunit

Zitieren

APA

Ngo, H., Niller, E., Schmitz, E., Kotter, E., Windfuhr-Blum, M., Neubauer, C., Palacios, A. L., Bamberg, F., Neubauer, J., Weiss, J., & Wilpert, C. (2026). Performance of a Screening Mammography AI Algorithm Repurposed for Symptomatic Mammography in a Tertiary Outpatient Clinic. Diagnostics, 16(7), 984. https://doi.org/10.3390/diagnostics16070984

Vancouver

Ngo H, Niller E, Schmitz E, Kotter E, Windfuhr-Blum M, Neubauer C, et al. Performance of a Screening Mammography AI Algorithm Repurposed for Symptomatic Mammography in a Tertiary Outpatient Clinic. Diagnostics. 2026;16(7):984. doi:10.3390/diagnostics16070984

BibTeX
@article{Ngo2026,
  author = {Ngo, Helen and Niller, Eric and Schmitz, Eric and Kotter, Elmar and Windfuhr-Blum, Marisa and Neubauer, Claudia and Palacios, Ana-Luisa and Bamberg, Fabian and Neubauer, Jakob and Weiss, Jakob and Wilpert, Caroline},
  title = {Performance of a Screening Mammography AI Algorithm Repurposed for Symptomatic Mammography in a Tertiary Outpatient Clinic},
  journal = {Diagnostics},
  year = {2026},
  volume = {16},
  number = {7},
  pages = {984},
  doi = {10.3390/diagnostics16070984}
}
Kostenlos starten →