Human In Vivo Cardiac Magnetic Resonance Imaging at 7 T: Feasibility, Applications, and Current Limitations—A Systematic Review
Diagnostics · 22.03.2026 · 16 (6) · 937
DOI: 10.3390/diagnostics16060937 ↗Kurz gefasst
Übersichtsarbeit in der Zeitschrift Diagnostics, online erschienen am 22.03.2026. Verfasst von Arosh S Perera Molligoda Arachchige und 7 weiteren Autor:innen. Beteiligte Einrichtung: Università degli Studi di Milano, Via Festa del Perdono 7, 20122 Milan, Italy. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: Cardiac magnetic resonance (CMR) imaging at 7 Tesla provides a substantially higher intrinsic signal-to-noise ratio compared with conventional 1.5 T and 3 T systems, potentially enabling higher spatial resolution, improved tissue contrast, and advanced metabolic imaging. However, clinical translation remains limited by technical challenges associated with ultra-high-field operation. This systematic review aimed to synthesize current human in vivo evidence on the feasibility, applications, and methodological limitations of 7-T cardiovascular MRI.
Methods
A PRISMA-guided systematic search of PubMed, Cochrane Library, Web of Science, and Scopus was conducted from database inception through January 2025. Studies reporting human in vivo cardiovascular MRI at 7 Tesla were included. Data regarding study design, sample characteristics, imaging applications, feasibility, quantitative findings, and reported limitations were extracted and qualitatively synthesized.
Results
Sixty-five studies met inclusion criteria, predominantly small prospective cohorts (mean sample size = 13), largely involving healthy volunteers. Across diverse applications—including coronary MR angiography, cine imaging, valvular assessment, vascular imaging, flow quantification, myocardial tissue characterization, and multinuclear (31P, 23Na, 39K) imaging—7-T CMR was consistently feasible and capable of producing high-quality images. Quantitative ventricular and vascular measurements were generally concordant with lower field strengths. Incremental benefits were most apparent in high-resolution structural imaging and metabolic applications, whereas routine functional and flow assessments showed limited additional advantages. No serious adverse events were reported.
Conclusions
Human cardiovascular MRI at 7 Tesla represents a technically feasible research and early translational platform with selective advantages over established field strengths. Further advances in radiofrequency technology, protocol harmonization, and larger disease-focused studies are required to clarify its potential clinical role.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Università degli Studi di Milano, Via Festa del Perdono 7, 20122 Milan, Italy.
- Faculty of Medicine & Surgery, Humanitas University, 20089 Rozzano, Italy.
- Department of Medical Biotechnology and Translational Medicine, International Medical School, Università degli Studi di Milano, 20054 Milan, Italy.
- Dipartimento di Scienze della Salute, Università degli Studi di Milano, 20054 Milan, Italy.
- IRCCS SYNLAB SDN, 80143 Naples, Italy.
ORCID-Profile: Arosh S Perera Molligoda Arachchige, Gabriel Amorim Moreira Alves, Oana-Georgiana Voicu, Erica Maffei, Filippo Cademartiri
Zitieren
APA
Perera Molligoda Arachchige, A. S., Amorim Moreira Alves, G., Zal, A., D'Acunto, G., Węglarz, M., Voicu, O. G., Maffei, E., & Cademartiri, F. (2026). Human In Vivo Cardiac Magnetic Resonance Imaging at 7 T: Feasibility, Applications, and Current Limitations—A Systematic Review. Diagnostics, 16(6), 937. https://doi.org/10.3390/diagnostics16060937
Vancouver
Perera Molligoda Arachchige AS, Amorim Moreira Alves G, Zal A, D'Acunto G, Węglarz M, Voicu OG, et al. Human In Vivo Cardiac Magnetic Resonance Imaging at 7 T: Feasibility, Applications, and Current Limitations—A Systematic Review. Diagnostics. 2026;16(6):937. doi:10.3390/diagnostics16060937
BibTeX
@article{PereraMolligodaArachchige2026,
author = {Perera Molligoda Arachchige, Arosh S and Amorim Moreira Alves, Gabriel and Zal, Ayça and D'Acunto, Giulia and Węglarz, Maciej and Voicu, Oana-Georgiana and Maffei, Erica and Cademartiri, Filippo},
title = {Human In Vivo Cardiac Magnetic Resonance Imaging at 7 T: Feasibility, Applications, and Current Limitations—A Systematic Review},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {6},
pages = {937},
doi = {10.3390/diagnostics16060937}
}
EUVIMED