Comparative Assessment of Proximal Humeral Bone Density Using CT Osteoabsorptiometry, Bone Microarchitecture Analysis, and a HU-Based Calibration Method: A CT and Micro-CT Study in Elderly Body Donors (65–86 Years)
Diagnostics · 03.03.2026 · 16 (5) · 756
DOI: 10.3390/diagnostics16050756 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 03.03.2026. Verfasst von Susanne Strasser und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Traumatology, Medical University Innsbruck, 6020 Innsbruck, Austria. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Local bone quality of the proximal humerus is a key determinant of fracture risk and implant stability in osteoporotic bone. Beyond established HU-based calibration, CT-osteoabsorptiometry (CT-OAM)-derived indices and microarchitecture-oriented workflows warrant systematic cross-modality evaluation.
Methods
Twelve proximal humeral heads from six body donors (age 65–86 years; bilateral specimens) were analyzed using paired clinical CT and high-resolution micro-CT. Bone quality was quantified by (i) a HU-calibrated cancellous vBMD method (Krappinger et al.), (ii) a CT-OAM-inspired workflow reporting an ROI-averaged mean-intensity index in arbitrary units (a.u.), and (iii) a calibrated Bone Microarchitecture Analysis (BMA) workflow in Analyze 15.0. Paired tests, linear regression, and repeated-measures ANOVA after z-standardization were applied.
Results
HU calibration yielded a mean trabecular vBMD of 114.37 ± 35.15 mg/cm3 on clinical CT. The BMA workflow produced higher CT-based values (207.37 ± 23.78 mg/cm3, p < 0.001) and markedly higher micro-CT values (469.34 ± 30.99 a.u.), indicating a systematic level shift between calibration frameworks. The CT-OAM index averaged 166.94 ± 40.12 a.u. on clinical CT and 455.89 ± 132.63 a.u. on micro-CT. Cross-modality agreement was very strong for CT-OAM (R2 = 0.888) and moderate for BMA (R2 = 0.502). After z-standardization, no significant differences were detected between the three CT-based approaches.
Conclusions
A CT-OAM-inspired ROI-mean index and a BMA-based workflow provide complementary, transferable readouts of proximal humeral bone quality across clinical CT and micro-CT, with stronger cross-modality rank consistency for CT-OAM. Absolute density values differ systematically between calibration frameworks and should not be interpreted as directly interchangeable. These approaches support opportunistic, site-specific bone quality assessment from routine CT, but require prospective validation against fixation-related outcomes and robust scanner-independent standardization.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Orthopaedics and Traumatology, Medical University Innsbruck, 6020 Innsbruck, Austria.
ORCID-Profile: Susanne Strasser, Johannes Dominikus Pallua
Zitieren
APA
Strasser, S., Adam, L., Kampik, L., Arora, R., & Pallua, J. D. (2026). Comparative Assessment of Proximal Humeral Bone Density Using CT Osteoabsorptiometry, Bone Microarchitecture Analysis, and a HU-Based Calibration Method: A CT and Micro-CT Study in Elderly Body Donors (65–86 Years). Diagnostics, 16(5), 756. https://doi.org/10.3390/diagnostics16050756
Vancouver
Strasser S, Adam L, Kampik L, Arora R, Pallua JD. Comparative Assessment of Proximal Humeral Bone Density Using CT Osteoabsorptiometry, Bone Microarchitecture Analysis, and a HU-Based Calibration Method: A CT and Micro-CT Study in Elderly Body Donors (65–86 Years). Diagnostics. 2026;16(5):756. doi:10.3390/diagnostics16050756
BibTeX
@article{Strasser2026,
author = {Strasser, Susanne and Adam, Lorenz and Kampik, Lukas and Arora, Rohit and Pallua, Johannes Dominikus},
title = {Comparative Assessment of Proximal Humeral Bone Density Using CT Osteoabsorptiometry, Bone Microarchitecture Analysis, and a HU-Based Calibration Method: A CT and Micro-CT Study in Elderly Body Donors (65–86 Years)},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {5},
pages = {756},
doi = {10.3390/diagnostics16050756}
}
EUVIMED