AI-Assisted Double-Headed Capsule Endoscopy: Multicentre Prospective Diagnostic Accuracy Study Across Small Bowel Indications
Diagnostics · 12.01.2026 · 16 (2) · 239
DOI: 10.3390/diagnostics16020239 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 12.01.2026. Verfasst von Kamran Mushtaq und 9 weiteren Autor:innen. Beteiligte Einrichtung: Southampton Interventional Endoscopy Unit, University Hospital Southampton NHS Foundation Trust, Southampton SO16 6YD, UK. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Aims: Double-headed capsule endoscopy enhances visualization and diagnostic yield in small bowel evaluation but increases reading time. This study aimed to assess the diagnostic performance of AI-assisted double-headed capsule endoscopy (MiroCam MC2000) across all small bowel indications and to compare its reading efficiency with the standard manual reading mode.
Methods
From May to December 2023, 242 consecutive patients (mean age 50.17 years, SD 18.3; 53% female) underwent small bowel capsule endoscopy at two UK centres for suspected Crohn’s disease (48.8%), iron-deficiency anemia (23.6%), bleeding (18.6%), or other (9%). Seven experienced readers reviewed videos in standard mode (blinded to clinical data), then AI-assisted (MiroCam AI Scan) methods were applied after de-identification/randomization. Two experts provided reference standards. No adverse events occurred.
Results
AI-assisted reading had sensitivity 95.3% (95% CI 90.1–98.3%) and specificity 96.5% (95% CI 91.3–99.0%) for diagnostic findings, vs. standard reading: 96.5% (95% CI 91.2–99.0%) and 85.3% (95% CI 78.0–90.9%). The positive findings rate was 83.6% vs. 80.2% (p = 0.040). Reading time decreased by 52% (38.1 vs. 18.26 min; p < 0.001).
Conclusions
AI-assisted reading offers high diagnostic accuracy, superior specificity and reduced reading times, supporting its adjunctive role with expert oversight. Registered: ERGO ID 82419.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Southampton Interventional Endoscopy Unit, University Hospital Southampton NHS Foundation Trust, Southampton SO16 6YD, UK.
- Department of Internal Medicine, Dongguk University Ilsan Hospital, Dongguk University College of Medicine, Goyang 10326, Republic of Korea.
- Department of Medicine, Gastroenterology and Endoscopy, Fondazione Policlinico Gemelli I.R.C.C.S., 00168 Rome, Italy.
- Gastroenterology Unit, Santa Maria delle Croci Hospital, 48121 Ravenna, Italy.
- Department of Clinical Research, University of Southern Denmark (SDU), 5230 Odense, Denmark.
- Department of Gastroenterology, Pomeranian Medical University, 70-111 Szczecin, Poland.
ORCID-Profile: Kamran Mushtaq, Yun Jeong Lim, Cristiano Spada, Alessandro Mussetto, Anastasios Koulaouzidis, Imdadur Rahman
Zitieren
APA
Mushtaq, K., Lim, Y. J., Spada, C., Mussetto, A., Koulaouzidis, A., Kaung, T., Borrow, D. M., Casadei, C., Patel, P., & Rahman, I. (2026). AI-Assisted Double-Headed Capsule Endoscopy: Multicentre Prospective Diagnostic Accuracy Study Across Small Bowel Indications. Diagnostics, 16(2), 239. https://doi.org/10.3390/diagnostics16020239
Vancouver
Mushtaq K, Lim YJ, Spada C, Mussetto A, Koulaouzidis A, Kaung T, et al. AI-Assisted Double-Headed Capsule Endoscopy: Multicentre Prospective Diagnostic Accuracy Study Across Small Bowel Indications. Diagnostics. 2026;16(2):239. doi:10.3390/diagnostics16020239
BibTeX
@article{Mushtaq2026,
author = {Mushtaq, Kamran and Lim, Yun Jeong and Spada, Cristiano and Mussetto, Alessandro and Koulaouzidis, Anastasios and Kaung, Thake and Borrow, Dean-Martin and Casadei, Cesare and Patel, Praful and Rahman, Imdadur},
title = {AI-Assisted Double-Headed Capsule Endoscopy: Multicentre Prospective Diagnostic Accuracy Study Across Small Bowel Indications},
journal = {Diagnostics},
year = {2026},
volume = {16},
number = {2},
pages = {239},
doi = {10.3390/diagnostics16020239}
}
EUVIMED