Contrast Volume Reduction in Oncologic Body Imaging Using Dual-Energy CT: A Comparison with Single-Energy CT
Diagnostics · 12.03.2025 · 15 (6) · 707
DOI: 10.3390/diagnostics15060707 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostics, online erschienen am 12.03.2025. Verfasst von Marianna Gulizia und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Interventional Radiology, Lausanne University Hospital, University of Lausanne (UNIL), 1011 Lausanne, Switzerland. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background/Objectives: To evaluate the feasibility of reducing contrast volume in oncologic body imaging using dual-energy CT (DECT) by (1) identifying the optimal virtual monochromatic imaging (VMI) reconstruction using DECT and (2) comparing DECT performed with reduced iodinated contrast media (ICM) volume to single-energy CT (SECT) performed with standard ICM volume.
Methods
In this retrospective study, we quantitatively and qualitatively compared the image quality of 35 thoracoabdominopelvic DECT across 9 different virtual monoenergetic image (VMI) levels (from 40 to 80 keV) using a reduced volume of ICM (0.3 gI/kg of body weight) to determine the optimal keV reconstruction level. Out of these 35 patients, 20 had previously performed SECT with standard ICM volume (0.3 gI/kg of body weight + 9 gI), enabling protocol comparison. The qualitative analysis included overall image quality, noise, and contrast enhancement by two radiologists. Quantitative analysis included contrast enhancement measurements, contrast-to-noise ratio, and signal-to-noise ratio of the liver parenchyma and the portal vein. ANOVA was used to identify the optimal VMI level reconstruction, while t-tests and paired t-tests were used to compare both protocols.
Results
VMI60 keV provided the highest overall image quality score. DECT with reduced ICM volume demonstrated higher contrast enhancement and lower noise than SECT with standard ICM volume (p < 0.001). No statistical difference was found in the overall image quality between the two protocols (p = 0.290).
Conclusions
VMI60 keV with reduced contrast volume provides higher contrast and lower noise than SECT at a standard contrast volume. DECT using a reduced ICM volume is the technique of choice for oncologic body CT.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology and Interventional Radiology, Lausanne University Hospital, University of Lausanne (UNIL), 1011 Lausanne, Switzerland.
- Faculty of Biology and Medicine (FBM), University of Lausanne (UNIL), 1015 Lausanne, Switzerland.
- Department of Nuclear Medicine, Lausanne University Hospital, University of Lausanne (UNIL), 1011 Lausanne, Switzerland.
ORCID-Profile: Guillaume Fahrni, Gibran Manasseh, Clarisse Dromain, Naik Vietti-Violi
Zitieren
APA
Gulizia, M., Viry, A., Jreige, M., Fahrni, G., Marro, Y., Manasseh, G., Chevallier, C., Dromain, C., & Vietti-Violi, N. (2025). Contrast Volume Reduction in Oncologic Body Imaging Using Dual-Energy CT: A Comparison with Single-Energy CT. Diagnostics, 15(6), 707. https://doi.org/10.3390/diagnostics15060707
Vancouver
Gulizia M, Viry A, Jreige M, Fahrni G, Marro Y, Manasseh G, et al. Contrast Volume Reduction in Oncologic Body Imaging Using Dual-Energy CT: A Comparison with Single-Energy CT. Diagnostics. 2025;15(6):707. doi:10.3390/diagnostics15060707
BibTeX
@article{Gulizia2025,
author = {Gulizia, Marianna and Viry, Anais and Jreige, Mario and Fahrni, Guillaume and Marro, Yannick and Manasseh, Gibran and Chevallier, Christine and Dromain, Clarisse and Vietti-Violi, Naik},
title = {Contrast Volume Reduction in Oncologic Body Imaging Using Dual-Energy CT: A Comparison with Single-Energy CT},
journal = {Diagnostics},
year = {2025},
volume = {15},
number = {6},
pages = {707},
doi = {10.3390/diagnostics15060707}
}
EUVIMED